Objective
- Make measurement samples from the new stocks of malachite green prepared
- Measure UV-Vis and fluorescence
Description
Procedure for stocks and fluorescence samples prepared by Dr.Hartings here
Data
- Stock solutions made by the group G: 0.3ppm, 1ppm, 3ppm, and 6ppm
- 0.30885 ppm (0.52711 g of 29.3 ppm Malachite green stock, water added to a total mass of 50.0 g)
- 1.00500 ppm (1.71704 g of 29.3 ppm Malachite green stock, water added to a total mass of 50.049 g)
- 3 ppm (5.1194 g of 29.3 ppm Malachite green stock, water added to a total mass of 50 g)
- 6 ppm (1.023 g of 29.3 ppm Malachite green stock, water added to a total mass of 10 g)
UV-Vis spectra for malachite green solutions
Graph of Concentration vs. Absorbance of UV-Vis spectra for malachite green solutions
Fluorescence of Rhodamine stock solutions
- Area under curve (Integration) TBD
The fluorescence standard curve of Rhodamine solution
Notes
- We measured UV-Vis, followed by fluorescence
- our concentrations(0.3ppm and 1ppm) were too dilute and now signal was picked up by fluorescence
- We then made two more solutions(3ppm and 6ppm) of higher concentrations and took fluorescence of them, but again, no signal was picked up
- For the 6ppm we ran out of 50 mL falcon tubes and used 15 mL ones instead and calculations were adjusted to make 10 g of the solution instead of 50.
- Dr. Hartings then prepared Rhodamine samples that were known to fluoresce
- Parameters Used:
- Start(nm): 565 End(nm): 700 Excitation(nm): 555
- Ex slit(nm):5 Em slit(nm): 50 Scan Speed(nm/min): 200
|