'''Caliper RQS VS Agilent RIN '''
'''Caliper RQS VS Agilent RIN '''


RNA metrics such as peak heights, peak areas, and concentration are used to determine an RQS (RNA Quality Score) number for each sample.  The Caliper RQS is a calculated score that rates the quality of RNA samples. The RQS has been validated to correlate well with Agilent’s RIN (RNA Integrity Number) and follows the same 0-10 scale rating. Results comparing RIN to RQS for the same samples run on both LabChip GX and Agilent’s Bioanalyzer 2100 typically show <10% deviation. The RQS is consistent over Caliper’s standard sensitivity RNA assay concentration range (25 to 250 ng/μL) with CV’s <20%.  For more information   
RNA metrics such as peak heights, peak areas, and concentration are used to determine an RQS (RNA Quality Score) number for each sample.  The Caliper RQS is a calculated score that rates the quality of RNA samples. The RQS has been validated to correlate well with Agilent’s RIN (RNA Integrity Number) and follows the same 0-10 scale rating. Results comparing RIN to RQS for the same samples run on both LabChip GX and Agilent’s Bioanalyzer 2100 typically show <10% deviation. The RQS is consistent over Caliper’s standard sensitivity RNA assay concentration range (25 to 250 ng/μL) with CV’s <20%.  For more information  http://www.caliperls.com/assets/022/8317.pdf