75
edits
Manlin Luo (talk | contribs) (New page: {{BioMicroCenter}} The Caliper LabChip GX is the most accurate and advanced nucleic acid separations system. The GX suite of instruments utilizes Caliper’s innovative microfluidics tec...) |
Manlin Luo (talk | contribs) No edit summary |
||
Line 7: | Line 7: | ||
*Flexible with 96 and 384 well compatibility | *Flexible with 96 and 384 well compatibility | ||
*Comprehensive and quantitative analysis with LabChip GX software | *Comprehensive and quantitative analysis with LabChip GX software | ||
=== [[BioMicroCenter:Caliper_INFO | Available Analyses]] === | |||
'''HiSens DNA ''' | |||
Sensitivity: | |||
Standard Workflow | |||
*5 pg/μL per fragment | |||
*100 pg/μL for smears | |||
Limited Sample Workflow | |||
*10 pg/μL per fragment | |||
*200 pg/μL for smears | |||
'''Total RNA ''' | |||
RNA metrics such as peak heights, peak areas, and concentration are used to determine an RQS (RNA Quality Score) number for each sample. The Caliper RQS is a calculated score that rates the quality of RNA samples. The RQS has been validated to correlate well with Agilent’s RIN (RNA Integrity Number) and follows the same 0-10 scale rating. Results comparing RIN to RQS for the same samples run on both LabChip GX and Agilent’s Bioanalyzer 2100 typically show <10% deviation. |
edits