User:Dustin Jacqmin/Notebook/Measurements Rotation/2013/02/07

From OpenWetWare
Jump to navigationJump to search
Project name Main project page

OSL Calibration Procedure

Created an OSL dose calibration for the following conditions:

  • Linac: C600
  • Energy: 6 MV
  • Depth: 1.5 cm (0.5 cm bolus, 1.0 cm SolidWater)
  • Backscatter: 10.0 cm SolidWater
  • Calibration Type: Multi-Dose
  • Dose Range: 0-300 cGy
  • Calibration OSL Dose Values: 0, 10, 50, 100, 200, 300 (cGy)
  • Standard OSL Dose Values: 50, 200 (cGy)

Irradiation Procedure