Lidstrom:Ultracentrifuge: Difference between revisions

From OpenWetWare
Jump to navigationJump to search
Line 15: Line 15:
** You probably won't be able to get any sample out of a collapsed tube.   
** You probably won't be able to get any sample out of a collapsed tube.   
** When they collapse it is very tricky to get them out.  '''Consult the rotor manual if you get tubes stuck.''' They recommend autoclaving the buckets upside down instead of trying to fish/pry the tubes out.
** When they collapse it is very tricky to get them out.  '''Consult the rotor manual if you get tubes stuck.''' They recommend autoclaving the buckets upside down instead of trying to fish/pry the tubes out.
=== How balanced does it need to be? ===
=== How balanced does it need to be? ===
* Tech support 5/2013 [[User:Janet B. Matsen|JM]] said that you can balance oppositely placed buckets by eye.  There is no need to spend the time weighing down to the nearest milligram.  The centrifuge will stop itself if it is off balance.  The auto shut-off feature is conservative enough that you cannot hurt the rotor by having too imbalanced of samples.  You can, however, waste your time if you need to rebalance and restart the ultracentrifuge.
* Tech support 5/2013 [[User:Janet B. Matsen|JM]] said that you can balance oppositely placed buckets by eye.  There is no need to spend the time weighing down to the nearest milligram.  The centrifuge will stop itself if it is off balance.  The auto shut-off feature is conservative enough that you cannot hurt the rotor by having too imbalanced of samples.  You can, however, waste your time if you need to rebalance and restart the ultracentrifuge.
=== The swinging buckets are numbered ===
* You can match the bucket number to the corresponding position on the rotor, or do something else deliberately.
** Janet's method: If my samples are ordered 1, 2, 3, 4 then I put them in bucket with their number. If the weights of 1 and 2 are balanced, and 3 and 4 are balanced I put the numbered buckets in a different position than their caps say.  It would be 1, 3, 2, 4 in the 1st, 2nd, 3rd, and 4th position.  I also make a sharpie label on each disposable tube to eliminate the possibility of mixing up samples in this process. 
=== When it wont' try to start ===
=== When it wont' try to start ===
* If the ultracentrifuge won't start and won't even try to start, open the door, lift the rotor up, place it back down, let it cool back down to your desired temp, and hold start.
* If the ultracentrifuge won't start and won't even try to start, open the door, lift the rotor up, place it back down, let it cool back down to your desired temp, and hold start.