Godin:Protocols/Glow Research Auto Glow
From OpenWetWare
Revision as of 19:50, 24 June 2010 by
Noam S. Lightstone
(
talk
|
contribs
)
(
→
Calibration Procedure
)
(
diff
)
← Older revision
|
Latest revision
(
diff
) |
Newer revision →
(
diff
)
Jump to navigation
Jump to search
Home
Calendar
Contact
Lab Members
Protocols
Publications
Research
Orders
Auto Glow
Glow Research
Auto Glow
Contents
1
Operation Procedure
1.1
Start
1.2
Stop
2
Calibration Procedure
Operation Procedure
Start
Turn on the vacuum pump at the pump
Press
AC ON
Insert sample in chamber
Place chamber lid
Close the door properly (The RF cannot be activated unless the door is closed)
Press
Vacuum
Press
Tuning Auto/Man
(Light turns on)
Set the timer as desired (30 s recommended)
Turn on gas flow
Press
Process Start
; the process will begin automatically once the pressure falls below 0.5 Torr.
You may press
Abort Process
if you want to terminate manually
Stop
Push
Vacuum
to vent to atmosphere
Stop flow
Remove sample
Turn off the machine
Turn off the vacuum pump
Calibration Procedure
Turn on the vacuum pump at the pump
Press
AC ON
Switch
Selector
to RF Set Point
Set RF to Zero
Switch
Selector
to Pressure
Press
Vacuum
Once the pressure falls to to 0.5 Torr, press
Process Start
Turn on gas flow by flipping GAS 1 toggle switch up
Let the pressure reading stabilize
Set the pressure to approximately 1 Torr by adjusting the gas flow. The system can operation in the range of 0.5 to 1.2 Torr
Press
Process Abort
Press
Vacuum
Close
Gas Flow
Navigation menu
Page actions
Page
Talk
Read
View source
History
Refresh
Page actions
Page
Talk
More
Tools
Personal tools
Log in
Request account
Navigation
Main Page
Recent changes
Help
Contact OWW
Add a Lab Notebook
Special pages
Search
research
Materials
Protocols
Resources
Tools
What links here
Related changes
Printable version
Permanent link
Page information
Cite this page
Browse properties