NANOpure Diamond Life Science (UV/UF)

scheduled maintenance

  • 12/9/06 - replace 0.2 micron Filter
  • 6/9/07 - replace UV Lamp
  • 6/9/07 - System Sanitization

maintenance requirements

  • Operation Manual
  • Cartridge Pack - replace when purity level is too low, organic levels are too high, or 0.2 micron filter clogs rapidly after installation
  • Ultrafilter - replace when sanitization of system is no longer effective (pyrogen levels too high)
  • System Sanitization - roughly once every 6 months (varies with use)
  • UV lamp - replace every 5-6 months
  • 0.2 micron filter - replace in any of the following conditions:
    • every 30 days
    • flow rate reduced
    • bacteria break through
  • Fuses - replace if blown

maintenance history

  • Kelsic 18:31, 9 November 2006 (EST)
    1. replaced UV lamp
    2. replaced Cartridge Pack
    3. replaced Ultrafilter
    4. rinsed and sanitized system
    5. replaced 0.2 micron filter