Mobile Phases
From OpenWetWare
Jump to navigationJump to search
Mobile Phases
1.Always use fresh LC-MS grade solvents
Thermo Fisher Scientific recommends replacing your solvents at least once every two weeks. Since we use lightly the Orbitrap, we change mobile phases A & B once every month.
Mobile Phase A = Loading Pump Buffer:
for 150mL | for 200mL | |
---|---|---|
100% H20 | 150mL | 200mL |
0.1% FA | 150 uL | 200 uL |
Mobile Phase B
for 150mL | for 200mL | |
---|---|---|
20% H20 | 30mL | 40mL |
80% ACN | 120 uL | 160 uL |
0.08% FA | 120uL | 160uL |
Wash solution
for 500mL | |
---|---|
10% Methanol | 500mL |
0.05% FA | 250uL |
Note. Label each the bottle with component
2. Sonicate for 45 minutes each
3. After changing the new NC pump to 0.100ul/min