SWTH Rhodopsin Amplified DNA
- PCR done by Uma 11/8/2016, Will Richardson Samples + Dan Inserillo Samples
- Band in the PCR -ve well indicates DNA contamination
- PCR done by Uma 11/14/2016, Will Richardson Samples
- Test for contamination.
- First 4 samples run using old reagents used in previous PCR.
- Second 4 samples, changed water, primer mix, and Qiagen Master Mix
- Contamination persists indicated by band in the PCR -ve well
- PCR done by Winnie 11/15/2016, Will Richardson Samples
- Test for contamination
- All new reagents used
- Contamination persists indicated by band in the PCR -ve well
- PCR done by Winnie and Uma 11/16/2016, Will Richardson Samples
- Test for contamination
- Well 2 and 3 represent the repeated comparison of old reagents vs new reagents PCR -ves only, keeping the Qiagen Master Mix the same
- Well 5 - 9 represent PCR done with reagents from the Pre-PCR room and with the Pre-PCR room pipettes
- PCR preparation in the lab (Pre-PCR bench) still indicate contamination, however PCR preparation in the Pre-PCR room shows no contamination
- However, extract negative shows a band, so there is still the possibility of contamination during DNA extraction
|