HighPoint/CannonLab:Fluorescence Spectrometry: Difference between revisions
From OpenWetWare
Jump to navigationJump to search
(New page: {{CannonLabMenu}} ==Using the Fluorometer== ===Warnings=== # Open the sample chamber to close the shutters and protect the detector if the signal exceeds the maximum (~9). # Turn off th...) |
|||
Line 16: | Line 16: | ||
# Turn ON the spectrometer | # Turn ON the spectrometer | ||
#* look for the <span style="color:#FF0000">red Lamp Monitor LED</span> on front of the instrument to turn on | #* look for the <span style="color:#FF0000">red Lamp Monitor LED</span> on front of the instrument to turn on | ||
#* wait for the LED on the right side of the instrument to change from ORANGE to GREEN | #* wait for the LED on the right side of the instrument to change from <span style="color:orange">ORANGE</span> to <span style="color:green">GREEN</span> | ||
# Turn on the computer | # Turn on the computer | ||
#* user "lab", password "lab" | #* user "lab", password "lab" |
Revision as of 08:26, 6 February 2012
<owwmenu font="arial, helvetica, sans-serif" bold="1"
color="white" bgcolor="indigo" hovercolor="white" bghovercolor="gray" topfontsize="10" fontSize="8" pagewidth="750" image="CannonLabIMG01.JPG" lab="MGSC/CannonLab"> Home=Lab-Home Lab Members=#, People=People, Contact=Contact Information, Associates=Collaborators and Former Members Research=#,Protein Membrane=Protein Membrane Interactions, Osmolytes=Osmolyte Effects Protocols=#, Vesicles=Lipid Vesicle Preparation, Solutions=Buffer and Osmolyte Solutions, Fluorescence=Fluorescence Spectrometry, UV-Vis=UV-Vis Spectrometry, Pipetting=Pipetting Exercise Notebooks=#, Master List=Notebook, Dr Cannon=Notebook/Jonathan Cannon Links=#, Courses=Course Pages
</owwmenu>
Using the Fluorometer
Warnings
- Open the sample chamber to close the shutters and protect the detector if the signal exceeds the maximum (~9).
- Turn off the lamp before shutting down the instrument.
- Be careful with the cuvettes. You will break one if you work enough, but try to form careful habits.
Using the SLM Aminco-Bowman Luminescence Spectrometer
Startup
- Make sure the computer is OFF
- Turn ON the "Lamp Enable" switch on the left side of the spectrometer
- Turn ON the spectrometer
- look for the red Lamp Monitor LED on front of the instrument to turn on
- wait for the LED on the right side of the instrument to change from ORANGE to GREEN
- Turn on the computer
- user "lab", password "lab"
- Open AB2 program
- ignore the error message
- wait for "Initializing Instrument" to finish
- follow the remainder of instructions beginning on p. 4-1 of the manual
Shutdown
- 20 Minutes before turning off the spectrometer turn OFF the "Lamp Enable" switch
- This is crucial to extend the life of the expensive arc lamp
- Turn off the computer at any time
- Clean the cuvettes and return them to the rack behind the computer
- These are the other expensive and breakable item
- They are placed behind the computer when not in use to minimize chances of accidental damage