Beauchamp:Lab Notebook: Difference between revisions

From OpenWetWare
Jump to navigationJump to search
Line 11: Line 11:
Typical Protocol
Typical Protocol


Monday - electrode implantation
<p>Monday - electrode implantation<br>
1. check batteries on BAK stimulator - change if reading lower than 115V (reads ~ 135V with brand new complement of batteries)
1. check batteries on BAK stimulator - change if reading lower than 115V (reads ~ 135V with brand new complement of batteries)
2. check batteries on patient interface - change if reading lower than 18V (reads ~ 27V with brand new complement of batteries)
2. check batteries on patient interface - change if reading lower than 18V (reads ~ 27V with brand new complement of batteries)
Line 19: Line 19:
6. merge 2nd structural MRI from previous week with CT on Stealth
6. merge 2nd structural MRI from previous week with CT on Stealth
7.
7.
</p>


==Orders==
==Orders==
staples
staples

Revision as of 04:31, 11 October 2007

Brain picture
Beauchamp Lab




MRI

  1. UT-Only Wiki

Electrophysiology

Typical Protocol

Monday - electrode implantation
1. check batteries on BAK stimulator - change if reading lower than 115V (reads ~ 135V with brand new complement of batteries) 2. check batteries on patient interface - change if reading lower than 18V (reads ~ 27V with brand new complement of batteries) 3. set up cabling for ITC to record from 16 channels (unplug channel 1 to test stimulation circuit) 4. check TTL pulse for gate and monitor out with stimulation circuit online 5. pick up CT scan from radiology library (signed HIPAA release required) after surgery (~11AM) 6. merge 2nd structural MRI from previous week with CT on Stealth 7.

Orders

staples