<!-- In your own words, write an easy-to-comprehend list of the steps your group took to set up the PCR reaction in the PCR reaction tubes. End with placing the tubes in to the thermal cycler. Do not copy-paste the instructions from the Workbook. That will be considered plagiarism. --> | <!-- In your own words, write an easy-to-comprehend list of the steps your group took to set up the PCR reaction in the PCR reaction tubes. End with placing the tubes in to the thermal cycler. Do not copy-paste the instructions from the Workbook. That will be considered plagiarism. --> | ||
# Step 1 | # Step 1 | ||
* Obtain all list materials above | |||
# Step 2 | # Step 2 | ||
# Step 3... | *The PCR tubes will need to be in strips of four, cut the initial strip of eight into a strip of four; OpenPCR machine only accepts strips of four or less. | ||
# Step 3 | |||
*WARNING label the tube with a TUBE LABEL, and ONLY WRITE on the LABELS; DO NOT WRITE on the TUBES. | |||
# Step 4 | |||
*Once tubes are correctly set-up place them in the rack | |||
#Step 5 | |||
*Label one of the tube as Positive Control | |||
#Step 6 | |||
*Label an additional tube as Negative Control | |||
#Step 7 | |||
*Set the pipette to transfer 50-MircoLiters | |||
#Step 8 | |||
*Attach a disposable pipette tip to the pipette | |||
#Step 9 | |||
*Using the pipette draw up 50-MircoLiters of PCR reaction mixture, and dispense the PCR mixture into the Positive Control tube. | |||
#Step 10 | |||
*Discard the disposable tip in the cup, after each use. Disposable tips are SINGLE USE ONLY, only one disposable tip per draw. i.e. draw-up PCR reaction mixture ten times it will require ten disposable tips. AVOID CONTAMINATION of the samples | |||
#Step 11 | |||
*Using the pipette, with a new disposable tip (single use only), draw up 50-MircoLiters of the Positive Control DNA/Primer Mix, and transfer it into the Positive Control tube. | |||
#Step 12 | |||
*Close the PCR tubes for Positive Control TIGHTLY | |||
#Step 13 | |||
*repeat steps 8-12 for NEGATIVE CONTROL now | |||
#Step 14 | |||
*Bring you tubes to TA at the PCR Machine, and with oversight set the PCR Machine as listed in OpenPCR Program. | |||
#Step 15 | |||
*With the APPROVAL of the TA place you tubes into the PCR Machine, and record the location of your tubes within the PCR Machine i.e. row & column. | |||
'''OpenPCR program''' | '''OpenPCR program''' |
Revision as of 23:33, 31 March 2015
BME 100 Spring 2015 | Home People Lab Write-Up 1 | Lab Write-Up 2 | Lab Write-Up 3 Lab Write-Up 4 | Lab Write-Up 5 | Lab Write-Up 6 Course Logistics For Instructors Photos Wiki Editing Help | |||||||||||||||||||||||||||||||
OUR TEAMLAB 4 WRITE-UPProtocolMaterials
PCR Reaction Sample List
OpenPCR program
Denature at 95°C for 30 seconds, Anneal at 57°C for 30 seconds, and *Extend at 72°C for 30 seconds
Research and DevelopmentPCR - The Underlying Technology
|