Magni:Protocols/tecan
From OpenWetWare
Infinite F200 readings
The microplate reader can be used for static or dynamic readings of bacteria cultures growth and fluorescent proteins.
Static readings
- Grow bacteria under the desired conditions.
- Transfer 200 ul of culture in a well of a 96-well microplate.
- Measure with or without lid.
- Always measure a "blank" that can be subtracted to the other raw measurements (blank for absorbance is sterile media; blank for absorbance is non-fluorescent culture grown to an OD600 comparable to the fluorescent cultures).
Dynamic readings
- Grow bacteria under the desired conditions.
- Transfer 200 ul of culture in a well of a 96-well microplate, excluding the 36 wells of the "frame".
- Dispense deionized water in all the unused wells.
- Add inducers to the cultures when required.
- Start a kinetic cycle of 18 hours with the i-control software (Tecan): 37degC during all the experiment, linear shaking (15 s, 3 mm amplitude), 5 s wait time (also 10 s is ok), absorbance reading (default parameters), fluorescence readings (gain=50 for medium and high copy plasmids, gain=50 or 80 for low copy plasmids, gain=80 or 100 for single-copy integrants), sampling time=5 min. The specified fluorescence gain factors are only guidelines and can be modified according to promoter/RBS strength, strain, growth medium and cell density. RFP filters are 535/620nm, GFP filters are 485/540nm.
- Always include a "blank" for absorbance and fluorescence, as described before.
- The microplate should be covered with lid for dynamic readings >5-6 hours.
List of filters
- for fluorescence (nm/bandwidth):
- 620/20
- 560/20
- 540/25
- 535/25
- 485/20 (we have three of this)
- 430/35
- for absorbance (nm):
- 600
- 340
- Commonly used filter pairs for fluorescent proteins (EX=excitation filter, EM=emission filter):
- CFP EX:430 EM:485
- GFP EX:485 EM:540
- YFP EX:485 EM:560
- RFP EX:535 EM:620