Lidstrom:Sorvall Centrifuge: Difference between revisions
From OpenWetWare
Jump to navigationJump to search
(→Rotors) |
(→Rotors) |
||
(One intermediate revision by the same user not shown) | |||
Line 14: | Line 14: | ||
**** Actually, Amanda might have autoclaved them before. | **** Actually, Amanda might have autoclaved them before. | ||
***Can't write on body; can only write on cap. Sandy says stickers don't stick. | ***Can't write on body; can only write on cap. Sandy says stickers don't stick. | ||
* GS-3. | * [http://www.thermoscientific.fr/eThermo/CMA/PDFs/Various/File_582.pdf GS-3]. | ||
* | ** For bigger bottles. | ||
* ???. | |||
** Holds eppendorf (1.5mL) tubes | |||
* For reference, the benchtop 50mL centrifuges (e.g. [https://static.thermoscientific.com/images/D21714~.pdf ST-15R] with a [https://extranet.fisher.co.uk/webfiles/fr/Pjointes/Documentation/SOR007_EN%20SORVALL%20ST16&ST40%20CENTRIFUGE.pdf TX-400] swinging bucket rotor) we have spin at 4696 RCF. | * For reference, the benchtop 50mL centrifuges (e.g. [https://static.thermoscientific.com/images/D21714~.pdf ST-15R] with a [https://extranet.fisher.co.uk/webfiles/fr/Pjointes/Documentation/SOR007_EN%20SORVALL%20ST16&ST40%20CENTRIFUGE.pdf TX-400] swinging bucket rotor) we have spin at 4696 RCF. | ||
Revision as of 17:03, 11 June 2014
Back to Protocols
Picture
Rotors
Rotors we have:
- SS-34. Holds 8*50 mL, spins at 20,000 RPM/36,900 RCF
- Tubes:
- Can't be autoclaved.
- Actually, Amanda might have autoclaved them before.
- Can't write on body; can only write on cap. Sandy says stickers don't stick.
- Can't be autoclaved.
- Tubes:
- GS-3.
- For bigger bottles.
- ???.
- Holds eppendorf (1.5mL) tubes
- For reference, the benchtop 50mL centrifuges (e.g. ST-15R with a TX-400 swinging bucket rotor) we have spin at 4696 RCF.
- There are stickers with maximum speeds. These were set by Katya, a Lidstrom Lab manager, circa 2011. It is unknown what the motivation was for keeping the rotors below maximum speed.
Instructions
Example protocol for AM1:
Janet's SIP3-4 centrifugation:
- Chill to 4oC with SS-34 rotor inside
- Fill correct tubes
- Spin for ___ min at ___ RCF at ___oC
Centrifugation Math & Basics
- Sigma summary:
- sedimentation rate is proportional to RFC, and proportional to the square of the particle diameter (cell diameter).