The fluorescence level measured by flow cytometry is in arbitrary units and entirely dependent on the instrument settings used in taking the measurement. Even when consistent settings are used, variability in instrument performance makes comparison between data sets acquired on different days uncertain unless the instrument is calibrated.  
The fluorescence level measured by flow cytometry is in arbitrary units and entirely dependent on the instrument settings used in taking the measurement. Even when consistent settings are used, variability in instrument performance makes comparison between data sets acquired on different days uncertain unless the instrument is calibrated.  


Running a sample of calibration beads as a control in flow experiments allows normalization of multiple data sets for comparison (even if acquired with different pmt settings or on different cytometers). Additionally, the level of bead fluorescence can be experimentally correlated to absolute units, such as Molecules of Equivalent GFP (MEG). This allows users to approximate copies per cell of GFP from flow data, based on [[MEG|earlier quantitative western blot characterization]].  
Running a sample of calibration beads as a control in flow experiments allows normalization of multiple data sets for comparison (even if acquired with different pmt settings or on different cytometers). Additionally, the level of bead fluorescence can be experimentally correlated to absolute units, such as Molecules of Equivalent GFP (MEG). This allows users to approximate copies per cell of GFP from flow data, based on earlier quantitative western blot characterization. See a first pass at MEG calibration [[MEG|here]].


= Beads =
= Beads =
Line 17: Line 17:
6.0- 6.4 um beads. (Size recommended by Glenn Paradis at the MIT flow facility. Not sure if there’s a strong rationale for this.)
6.0- 6.4 um beads. (Size recommended by Glenn Paradis at the MIT flow facility. Not sure if there’s a strong rationale for this.)


The cost of these beads is $235.00 per 5 mL, or < $0.05 per experiment, particularly if you store and reuse diluted beads, as recommended.  
The cost of these beads is $235.00 per 5 mL, or < $0.05 per experiment, particularly if you store and reuse diluted beads, as recommended.
 
Note: As a test, beads were diluted on 12.15.05 and stored at -4C in an opaque container. The same stock of beads were freshly diluted on 01.16.06 and both tubes were compared by FACS. To the accuracy of the instrument, the measurements were identical (+/- 2%). -[[User:Jkm|Jkm]]
 
==FlowJo Calibration==
 
In FlowJo, select the bead sample to calibrate against. Then choose Platform->Define Calibration. Select the parameter to calibrate, and use the beads as your reference sample. I ([[User:Jkm|Jkm]]) use as reference values:<br>
Peak  GFP    RFP
1.    10.3    31.1
2.    30.2    70.4
3.    83.7  224
4.  315    885
5.  1092    2922
 
These numbers are arbitrary - they simply are the values that I measured for one specific date. Every other measurement, then, is calibrated against that day.
 
Occasionally, these peaks are difficult to distinguish (if the setting are extremely high or low, peaks will not be apparent). It can become a judgment call to decide which measured peak corresponds to which calibration peak.


= Uncertainty in Calibration =
= Uncertainty in Calibration =


*Qualitative observations from Josh, via cmc (Josh, feel free to elaborate on this or correct me if I’m wrong. ~ [[User:Cconboy|cmc]] 11:42, 10 Nov 2005 (EST))  
*Qualitative observations from Josh (all numbers rough)
**Uncertainty <5% when calibrating data sets from the same instrument with the same settings.  
**Uncertainty <5% when calibrating data sets from the same instrument with the same settings on different days.
**Uncertainty ~10% when calibrating data acquired on the same instrument with variable pmt settings.
***Variation of <5% on the bead readings day to day, same machine and PMT settings.
**Uncertainty ~10% when calibrating data acquired on the same instrument with variable pmt settings on the same day.
***On one day, one machine, run samples and beads, change PMT settings, run samples and beads, etc.
**Uncertainty ~20% when calibrating between different pmt settings if only using one calibration peak.
**Uncertainty ~20% when calibrating between different pmt settings if only using one calibration peak.
***Same as above, only using a single point calibration.
**Uncertainty when calibrating different machines hasn’t been evaluated.  
**Uncertainty when calibrating different machines hasn’t been evaluated.  
*Bead quality issues:
*Bead quality issues:
**Bead lifetime. (“very stable under proper storage conditions” means…?)
**Bead lifetime. (“very stable under proper storage conditions” means…?)
**Lot to lot bead variability. Could retain one master lot to calibrate all new lots against. This comes back to bead lifetime eventually. Certainly any new lot of beads should be run against the previous lot.  
**Lot to lot bead variability. Could retain one master lot to calibrate all new lots against. This comes back to bead lifetime eventually. Certainly any new lot of beads should be run against the previous lot.


= Resources =
= Resources =
Line 47: Line 66:


[[Caitlin Conboy]], [[Josh Michener]], [[Francois St.Pierre]], etc.
[[Caitlin Conboy]], [[Josh Michener]], [[Francois St.Pierre]], etc.
[[Category:Protocol]] [[Category:Flow cytometry]]

Latest revision as of 11:36, 9 May 2007

Background

The fluorescence level measured by flow cytometry is in arbitrary units and entirely dependent on the instrument settings used in taking the measurement. Even when consistent settings are used, variability in instrument performance makes comparison between data sets acquired on different days uncertain unless the instrument is calibrated.

Running a sample of calibration beads as a control in flow experiments allows normalization of multiple data sets for comparison (even if acquired with different pmt settings or on different cytometers). Additionally, the level of bead fluorescence can be experimentally correlated to absolute units, such as Molecules of Equivalent GFP (MEG). This allows users to approximate copies per cell of GFP from flow data, based on earlier quantitative western blot characterization. See a first pass at MEG calibration here.

Beads

Note: Any beads can be used. As long as they’re run against The Default Bead the data can be normalized. As a general rule, beads with multiple intensity peaks will give better calibration (see uncertainty below). Beads with a rainbow of fluorophores are most versatile.

The Default Bead

SPHERO Rainbow Calibration Particles from Spherotech:

Cat.#: RCP-60-5

6.0- 6.4 um beads. (Size recommended by Glenn Paradis at the MIT flow facility. Not sure if there’s a strong rationale for this.)

The cost of these beads is $235.00 per 5 mL, or < $0.05 per experiment, particularly if you store and reuse diluted beads, as recommended.

Note: As a test, beads were diluted on 12.15.05 and stored at -4C in an opaque container. The same stock of beads were freshly diluted on 01.16.06 and both tubes were compared by FACS. To the accuracy of the instrument, the measurements were identical (+/- 2%). -Jkm

FlowJo Calibration

In FlowJo, select the bead sample to calibrate against. Then choose Platform->Define Calibration. Select the parameter to calibrate, and use the beads as your reference sample. I (Jkm) use as reference values:

Peak  GFP    RFP
1.    10.3    31.1
2.    30.2    70.4
3.    83.7   224
4.   315     885
5.  1092    2922

These numbers are arbitrary - they simply are the values that I measured for one specific date. Every other measurement, then, is calibrated against that day.

Occasionally, these peaks are difficult to distinguish (if the setting are extremely high or low, peaks will not be apparent). It can become a judgment call to decide which measured peak corresponds to which calibration peak.

Uncertainty in Calibration

  • Qualitative observations from Josh (all numbers rough)
    • Uncertainty <5% when calibrating data sets from the same instrument with the same settings on different days.
      • Variation of <5% on the bead readings day to day, same machine and PMT settings.
    • Uncertainty ~10% when calibrating data acquired on the same instrument with variable pmt settings on the same day.
      • On one day, one machine, run samples and beads, change PMT settings, run samples and beads, etc.
    • Uncertainty ~20% when calibrating between different pmt settings if only using one calibration peak.
      • Same as above, only using a single point calibration.
    • Uncertainty when calibrating different machines hasn’t been evaluated.
  • Bead quality issues:
    • Bead lifetime. (“very stable under proper storage conditions” means…?)
    • Lot to lot bead variability. Could retain one master lot to calibrate all new lots against. This comes back to bead lifetime eventually. Certainly any new lot of beads should be run against the previous lot.

Resources

Bead Suppliers:

Discussions:

OWW Contacts:

Caitlin Conboy, Josh Michener, Francois St.Pierre, etc.