OBJECTIVES
- Label and remove previous PVOH films from Teflon dishes
- Select two types of PVOH for viscosity testing and measurements
- Complete viscosity calculations and graphing
Viscosity Preparations and Calculations
Creating Stock Solutions:
- In 600mL beaker, add 200mL of H2O and ~ 2 grams PVOH (MW 22K) with a large magnetic stir bar
(Initial Mass = 2.0256g)
- Heat and stir solution for ~ 45 minutes at 100 - 120°C
- Over a period of ~ 30 minutes, add an additional ~ 5 grams of PVOH (MW 22K) to solution, stirring frequently to limit frothing
(Total Mass In Solution = 7.0158g) (Concentration = 638μM)
- When PVOH thoroughly dissolves, transfer solution to 500mL volumetric flask to cool to room temperature
- After cooling, add more water to the volumetric flask to bring solution to volume
* Repeat the entire process again; however use PVOH (MW 130K)
(Initial Mass = 2.0342g) (Total Mass In Solution = 7.0545g) (Concentration = 109μM)
Dilution Standards for Stock Solutions PVOH:
- In 5 separate 50mL volumetric flasks, select and create various dilute PVOH (MW 22K) solutions for viscosity testing
- After creating dilute samples, begin testing the viscosity rates by pouring one solution, ~ 15mL, into a viscometer
(Dilute Sample Concentrations PVOH (MW 22K) = 79.7μM, 159μM, 319μM, 478μM)
- Using vacuum pump, suck up a solution into the viscometer until it reaches the top line of the first bulb on viscometer
- Once the solution surpasses the second line on the first bulb, begin timing to measure viscosity rate of the solution
- Take three readings for each solution, then average the time trials
* Repeat the entire process again; however use PVOH (MW 130K) stock solution
(Dilute Sample Concentrations PVOH (MW 130,000) = 13.6μM, 27.1μM, 54.3μM, 81.4μM)
Viscosity Charts and Tables PVOH (MW 22K)
Viscometer Model 2189: (40°C / 100°F - 0.003966 and 100°C / 210°F - 0.003950)
Viscosity of pure H2O
Trial [ ]
|
Time [s]
|
1 |
278.69
|
2 |
283.09
|
3 |
280.24
|
Average |
241.01
|
Viscosity of 79.7μM
Trial [ ]
|
Time [s]
|
1 |
279.37
|
2 |
281.97
|
3 |
283.12
|
Average |
281.49
|
Viscosity of 159μM
Trial [ ]
|
Time [s]
|
1 |
246.42
|
2 |
450.31
|
3 |
251.86
|
Average |
316.20
|
Viscosity of 319μM
Trial [ ]
|
Time [s]
|
1 |
273.99
|
2 |
276.73
|
3 |
278.63
|
Average |
276.45
|
Viscosity of 478μM
Trial [ ]
|
Time [s]
|
1 |
272.81
|
2 |
278.39
|
3 |
280.43
|
Average |
277.21
|
Viscosity of 638μM
Trial [ ]
|
Time [s]
|
1 |
386.09
|
2 |
399.45
|
3 |
703.49
|
Average |
496.34
|
Viscosity Charts and Tables PVOH (MW 130K)
Viscometer Model 2189: (40°C / 100°F - 0.003966 and 100°C / 210°F - 0.003950)
(Use Viscosity Chart of Pure H2O Listed Above)
Viscosity of 13.6μM
Trial [ ]
|
Time [s]
|
1 |
263.59
|
2 |
263.12
|
3 |
263.01
|
Average |
263.24
|
Viscosity of 27.1μM
Trial [ ]
|
Time [s]
|
1 |
322.23
|
2 |
316.28
|
3 |
316.25
|
Average |
318.25
|
Viscosity of 54.3μM
Trial [ ]
|
Time [s]
|
1 |
400.68
|
2 |
398.66
|
3 |
398.73
|
Average |
399.36
|
Viscosity of 81.4μM
Trial [ ]
|
Time [s]
|
1 |
604.14
|
2 |
610.79
|
3 |
608.64
|
Average |
607.86
|
Viscosity of 109μM
Trial [ ]
|
Time [s]
|
1 |
956.96
|
2 |
957.38
|
3 |
957.46
|
Average |
960.60
|
Notes
- Pure H2O was used as a standard for establishing a viscosity rate for a pure substance
- Pure H2O was also used as a cleaning agent between trials
- During the first two trials for the PVOH (MW 22K), the viscometer was accidentally not cleaned between trials
(viscosity calculations effected for PVOH MW 22K)
|