Atomic Absorption Spectroscopy of Au/BSA
- A rack of Au/BSA solutions was recovered and used for the assigned Atomic Absorption Spectroscopy (AAS). AAS was used to strip the molecule to its elemental state.
- The entire solution set were observed to lack fibers even at high mole ratios where formation of fibers is typical. The date of which the solutions were made is unknown.
- The equipment used for AAS is by the make of Shimadzu AA-6200 Atomic Absorption Flame Emission.
- The sample cell used for this procedure is the ignited flame.
- A gold hollow cathode tube was selected for the procedure. The wavelength of the lamp was 242.8 nm.
- Two blanks were used; water and HCl. Water was chosen since it is part of the ternary mixture. The standards provided by User:Abigail E. Miller/Notebook/CHEM571 Lab Project/2012/10/08 Dr. Miller were made with HCl. In addition, HCl ionizes better and keeps the metal in solution.
|