User:Mary Mendoza/Notebook/CHEM572 Exp. Biological Chemistry II/2013/02/20

From OpenWetWare

(Difference between revisions)
Jump to: navigation, search
(ADA Kinetic Assay for obtaining the zero point)
(ADA Kinetic Assay for obtaining the zero point)
Line 18: Line 18:
* An article was provided by Dr. Hartings with the molar extinction coefficient, 1.53 x 10<sup>-4</sup> of adenosine at 260 nm.
* An article was provided by Dr. Hartings with the molar extinction coefficient, 1.53 x 10<sup>-4</sup> of adenosine at 260 nm.
* Thus, it was decided to run a full spectrum of adenosine at each given concentration specified above for accurate measurement of  the concentration of adenosine.
* Thus, it was decided to run a full spectrum of adenosine at each given concentration specified above for accurate measurement of  the concentration of adenosine.
 +
 +
==Data==
 +
 +
[[Image:Concen1.png]]
 +
 +
 +
[[Image:AveVelo1.png]]
 +
 +
[[Image:1adeno.png]]
 +
 +
[[Image:Lin1.png]]
<!-- ##### DO NOT edit below this line unless you know what you are doing. ##### -->
<!-- ##### DO NOT edit below this line unless you know what you are doing. ##### -->

Revision as of 13:34, 22 February 2013

Project name Main project page
Previous entry      Next entry

ADA Kinetic Assay for obtaining the zero point

  • The objective of this laboratory period is to conduct adenosine deaminase (ADA) kinetic assay runs for the new calculated concentrations.
  • UV 2550 Shimadzu spectrophotometer was baseline with 0.05 M sodium phosphate buffer.
  • The assays were prepared according to the data below.
  • After running the first trial, it was observed that the absorbance for 12.34 μM adenosine of trial 2 was superimposed over the 10.52 μM adenosine of trial 1.
  • It was suggested by Dr. Hartings to use Beer's Law for accurate measurement of the concentration of adenosine in solution.
  • An article was provided by Dr. Hartings with the molar extinction coefficient, 1.53 x 10-4 of adenosine at 260 nm.
  • Thus, it was decided to run a full spectrum of adenosine at each given concentration specified above for accurate measurement of the concentration of adenosine.

Data

Image:Concen1.png


Image:AveVelo1.png

Image:1adeno.png

Image:Lin1.png