User:Madeleine Y. Bee/Notebook/Single Molecule Fluorescence/2013/07/03: Difference between revisions

From OpenWetWare
Jump to navigationJump to search
Line 23: Line 23:
[[Image:2013_0703_DNA-MB_avg.PNG]]<br.>
[[Image:2013_0703_DNA-MB_avg.PNG]]<br.>
Three different concentrations, run three times, 500s each.  Samples were made using the same 1nM DNA and MB dilutions from 07/02/2013, allowed to hybridize at ~70C for 25 min then cooled for 20 min.  100pM DNA and 75pM DNA show expected trend: increase in y-intercept with decrease in concentration, but 50pM does not.<br.>
Three different concentrations, run three times, 500s each.  Samples were made using the same 1nM DNA and MB dilutions from 07/02/2013, allowed to hybridize at ~70C for 25 min then cooled for 20 min.  100pM DNA and 75pM DNA show expected trend: increase in y-intercept with decrease in concentration, but 50pM does not.<br.>
====Notes=====
====Notes====
The inconsistency of the 50pM DNA/40pM MB sample has been observed twice in a row, which may indicate a lower concentration limit around this concentration.  However, since both sets of samples that have shown this inconsistency were made from the same dilutions (made on 07/02/2013), new samples will be completely remade to see if this trend continues.  To look into this possibility further and for the construction of a calibration graph, 25pM DNA/20pM MB and 125pM DNA/100pM MB samples will be made for the next run.<br.>
The inconsistency of the 50pM DNA/40pM MB sample has been observed twice in a row, which may indicate a lower concentration detection limit around this concentration.  However, since both sets of samples that have shown this inconsistency were made from the same dilutions (made on 07/02/2013), new samples will be completely remade to see if this trend continues.  To look into this possibility further and for the construction of a calibration graph, 25pM DNA/20pM MB and 125pM DNA/100pM MB samples will be made for the next run.<br.>
PBS will also be run again, as the fluorescence observed was significantly higher than expected.  All samples have been diluted with PBS. <br.>
PBS will also be run again, as the fluorescence observed was significantly higher than expected.  All samples have been diluted with PBS. <br.>
Green fluorescent beads are always run first to align laser for all following samples.<br.>
Green fluorescent beads are always run first to align laser for all following samples.<br.>

Revision as of 13:25, 3 July 2013

FCS Measurements <html><img src="/images/9/94/Report.png" border="0" /></html> Main project page
<html><img src="/images/c/c3/Resultset_previous.png" border="0" /></html>Previous entry<html>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;</html>Next entry<html><img src="/images/5/5c/Resultset_next.png" border="0" /></html>

July 3, 2013

FCS Sample Prep

  • Green fluorescent beads
    • 10000x diluted, run for 90s
  • PBS
    • To check for background, run for 90s
  • DNA/MB
    • 100/80pM
    • 75/60pM
    • 50/40pM

FCS Data

<br.> Green fluorescent beads diluted 10000x were run three times, 90s each; used to properly align laser for other samples. Laser not adjusted after bead samples were run.<br.> <br.> PBS run twice, 90s each; to check for any possible background fluorescence. Average values show more fluorescence than expected, will be run again with next set of samples.<br.> <br.> Three different concentrations, run three times, 500s each. Samples were made using the same 1nM DNA and MB dilutions from 07/02/2013, allowed to hybridize at ~70C for 25 min then cooled for 20 min. 100pM DNA and 75pM DNA show expected trend: increase in y-intercept with decrease in concentration, but 50pM does not.<br.>

Notes

The inconsistency of the 50pM DNA/40pM MB sample has been observed twice in a row, which may indicate a lower concentration detection limit around this concentration. However, since both sets of samples that have shown this inconsistency were made from the same dilutions (made on 07/02/2013), new samples will be completely remade to see if this trend continues. To look into this possibility further and for the construction of a calibration graph, 25pM DNA/20pM MB and 125pM DNA/100pM MB samples will be made for the next run.<br.> PBS will also be run again, as the fluorescence observed was significantly higher than expected. All samples have been diluted with PBS. <br.> Green fluorescent beads are always run first to align laser for all following samples.<br.>