Line 37: Line 37:
**25pM/12.5pM: from 2nM/1.6nM dilutions
**25pM/12.5pM: from 2nM/1.6nM dilutions
===FCS Data: Calibration Curves===
===FCS Data: Calibration Curves===
-
[[Image:FCS_data_2013_0701_OD%2C_MB-DNA_calibrations.PNG]]
+
[[Image:FCS_data_2013_0701_OD%2C_MB-DNA_calibrations.PNG]]<br.>
 +
Green fluorescent beads were run for 90s to align laser.<br.>
 +
Oligo D samples were run for 300s.<br.>
 +
MB-DNA samples were heated to ~70C for 25 min, cooled for 20 min, and run at 500s.<br.>
====Notes====
====Notes====
*Use green fluorescent beads to align laser
*Use green fluorescent beads to align laser
*Calibration curves did not turn out as expected (linear trend)
*Calibration curves did not turn out as expected (linear trend)
*DNA, not MB, should be in excess for DNA-MB samples to ensure all MB is bound
*DNA, not MB, should be in excess for DNA-MB samples to ensure all MB is bound
-
*Focus on three concentrations, run each sample several times, average curves  
+
*Focus on three concentrations, run each sample several times, average curves based on best spectra taken today
**Oligo D: 150, 125, 100pM run three times at 500s per sample
**Oligo D: 150, 125, 100pM run three times at 500s per sample
**DNA/MB: 100pM DNA/80pM MB, 75/60, 50/40pM run three times at 500s per sample (increase to 700s per sample if spectra are noisey)
**DNA/MB: 100pM DNA/80pM MB, 75/60, 50/40pM run three times at 500s per sample (increase to 700s per sample if spectra are noisey)

Current revision

FCS Measurements Main project page
Previous entry      Next entry

July 1, 2013

FCS Sample Preparation

  • Green fluorescent beads
    • 1000x diluted: from 10x dilution
    • 10,000x diluted: from 10x dilution
  • Oligo D: 20,000pM → 1000pM → 150pM
    • 150pM: from 1000pM dilution
    • 125pM: from 150pM dilution
    • 100pM: from 150pM dilution
    • 80pM: from 150pM dilution
    • 60pM: from 150pM dilution
    • 50pM: from 150pM dilution
    • 40pM: from 150pM dilution
    • 30pM: from 150pM dilution
    • 20pM: from 150pM dilution
    • 10pM: from 150pM dilution
  • MB/DNA
    • 1nM/800pM: from 2nM/1.6nM dilutions
    • 900pM/720pM: from 50nM dilutions
    • 800pM/640pM: from 50nM dilutions
    • 700pM/560pM: from 2nM/1.6nM dilutions
    • 600pM/480pM: from 2nM/1.6nM dilutions
    • 500pM/400pM: from 2nM/1.6nM dilutions
    • 400pM/320pM: from 2nM/1.6nM dilutions
    • 300pM/240pM: from 2nM/1.6nM dilutions
    • 200pM/160pM: from 2nM/1.6nM dilutions
    • 100pM/80pM: from 2nM/1.6nM dilutions
    • 75pM/60pM: from 2nM/1.6nM dilutions
    • 50pM/40pM: from 2nM/1.6nM dilutions
    • 25pM/12.5pM: from 2nM/1.6nM dilutions

FCS Data: Calibration Curves

Image:FCS_data_2013_0701_OD,_MB-DNA_calibrations.PNG
Green fluorescent beads were run for 90s to align laser.
Oligo D samples were run for 300s.
MB-DNA samples were heated to ~70C for 25 min, cooled for 20 min, and run at 500s.

Notes

  • Use green fluorescent beads to align laser
  • Calibration curves did not turn out as expected (linear trend)
  • DNA, not MB, should be in excess for DNA-MB samples to ensure all MB is bound
  • Focus on three concentrations, run each sample several times, average curves based on best spectra taken today
    • Oligo D: 150, 125, 100pM run three times at 500s per sample
    • DNA/MB: 100pM DNA/80pM MB, 75/60, 50/40pM run three times at 500s per sample (increase to 700s per sample if spectra are noisey)