User:Linh N Le/Notebook/2009/06/17/How to use the Autoclave

From OpenWetWare
< User:Linh N Le‎ | Notebook‎ | 2009‎ | 06‎ | 17
Revision as of 14:48, 18 June 2009 by Linh N Le (talk | contribs) (→‎No drying)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigationJump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.

Initial Check

  • On the top of the autoclave is a round opening for DI water
  • The cap has a dipstick in it where you can check water levels
    • Note: do not over fill the autoclave. There is a metal spigot inside the filling area that the water level must not rise above

Preparing samples

  • Tubes should be placed in the round plastic screw cap containers
    • The cap should be placed upon the top, but not screwed in
  • Pipette tip boxes should be opened and their lids on, but not airtight
  • All samples must have the special masking tape on them
    • This tape is the manilla with white stripes on it
    • Make a small tab at one end so that the tape can be easily removed for reuse of the container.
  • Place the samples onto the tray of the autoclave, leaving space between items

Setup

  • After the autoclave has been filled with water and loaded, turn on the green switch to "START"
  • Turn the bottom-most dial clockwise to the "FILL WATER" position
  • Watch inside the autoclave bottom as water starts to fill it
  • When the water reaches the door (There is a small notch before the door) set the bottom dial to "STE"
  • Close door and tighten (not too tight because you may damage the gasket)
  • With the door closed, set the temperature gauge to 273F (121C)
  • Set the timer to 30mins (45mins if this is a cold start)
    • Note: the pressure meter must be between 30-32 psi for the temperature to actually be 273F. If it is not, adjust the temperature gauge accordingly

Liquids

  • Only set to 250F and use a non-perforated pan

Drying

  • After the timer is done, a small "ding" will be heard
  • Wait for the pressure to reach 0 psi
  • Open the door slightly
  • Set the bottom-most dial to "EXH + DRY"
  • Set the timer knob to 20-30mins
    • The DRY light should be on now
  • When the timer expires, change the bottom most knob to the "0" position
  • Open door and remove samples

No drying

  • After the timer dings set the bottom most knob to "EXH + DRY"
  • When the pressure gauge reaches 0 psi, turn the bottom most knob to the "0" position
  • Open door and remove samples

Liquids

  • Let the pressure reach 0
  • Switch bottom knob to "0"
  • Open door and enjoy

Make sure to switch the green switch to "STOP"