Quantification of glucose: Difference between revisions

From OpenWetWare
Jump to navigationJump to search
Line 1: Line 1:
==Introduction==
==Introduction==
==Method==
==Method==
SET-UP
===SET-UP===
# Rinse out Buffer container and fill to line with YSI 2357 Buffer solution
# Rinse out Buffer container and fill to line with YSI 2357 Buffer solution
# Hit [Menu} then {2-MeasParameter] then [2-CalMethod] then [1-One Station]
# Hit [Menu} then {2-MeasParameter] then [2-CalMethod] then [1-One Station]
Line 12: Line 12:




RUN SAMPLES
===RUN SAMPLES===
# Hit [Run] and wait for system to complete calibration
# Hit [Run] and wait for system to complete calibration
# When Ready screen appears, hit [Sample]
# When Ready screen appears, hit [Sample]
# Present sample to needle in Station S3 without touching needle to tube
# Present sample to needle in Station S3 without touching needle to tube
# Hit [Sample] and wait for needle to retract
# Hit [Sample] and wait for needle to retract
* Results will print out at the completion of run
# Results will print out at the completion of run
* To continue testing samples, hit [Sample] when Ready screen appears
# To continue testing samples, hit [Sample] when Ready screen appears
 


*Note: After 5-7 runs, system will automatically calibrate.
*Note: After 5-7 runs, system will automatically calibrate.

Revision as of 09:01, 5 July 2011

Introduction

Method

SET-UP

  1. Rinse out Buffer container and fill to line with YSI 2357 Buffer solution
  2. Hit [Menu} then {2-MeasParameter] then [2-CalMethod] then [1-One Station]
  3. Hit [Enter] twice to return to Measurement Parameter Set up Screen
  4. Hit [3-Black]
  5. Make sure Current BLACK probe chem is set to NONE. If it isn't, hit [1-Next chemistry] until NONE appears and hit [ENTER] to save
  6. Hit [4-White] and set Current WHITE probe chem to Dextrose
  7. Hit [Enter] 4 times to return to Measurement Parameter Set up Screen
  8. Hit [Menu] to return to home screen


RUN SAMPLES

  1. Hit [Run] and wait for system to complete calibration
  2. When Ready screen appears, hit [Sample]
  3. Present sample to needle in Station S3 without touching needle to tube
  4. Hit [Sample] and wait for needle to retract
  5. Results will print out at the completion of run
  6. To continue testing samples, hit [Sample] when Ready screen appears


  • Note: After 5-7 runs, system will automatically calibrate.