Lidstrom:Sorvall Centrifuge: Difference between revisions

From OpenWetWare
Jump to navigationJump to search
(10 intermediate revisions by the same user not shown)
Line 2: Line 2:




== Picture ==
== Background ==
[[image:Lidstrom_Sorvoll.png|thumb|upright=1.0|center|Lidstrom Sorvoll centrifuge]]
[[image:Lidstrom_Sorvoll.png|thumb|upright=1.0|center|Lidstrom Sorvoll centrifuge]]
We have a Sorvoll RC-5B centrifuge that is useful for spinning large volumes in 0.5L bottles or medium-sized (a few mL) volumes very quickly


== Rotors ==
== Rotors ==
* [http://www.geminibv.nl/labware/sorvall-gsa-rotor/sorvall-rotor-gsa-gs3-sa600-se12-sm24-ss34-manual.pdf Sorvall rotors]
[http://www.geminibv.nl/labware/sorvall-gsa-rotor/sorvall-rotor-gsa-gs3-sa600-se12-sm24-ss34-manual.pdf Sorvall rotors]
** Rotors we have:
 
*** SS-34.  Holds 8*50 mL, spins at 20,000 RPM/36,900 RCF
****Tubes: [[image:Lidstrom_SS-34 tubes.png|thumb|upright=0.3|center|SS-34 tubes]]
*****Can't be autoclaved. 
****** Actually, Amanda might have autoclaved them before. 
*****Can't write on body; can only write on cap.  Sandy says stickers don't stick.
*** GS-3.
*** ____.  Holds eppendorf (1.5mL) tubes
* For reference, the benchtop 50mL centrifuges (e.g. [https://static.thermoscientific.com/images/D21714~.pdf ST-15R] with a [https://extranet.fisher.co.uk/webfiles/fr/Pjointes/Documentation/SOR007_EN%20SORVALL%20ST16&ST40%20CENTRIFUGE.pdf TX-400] swinging bucket rotor) we have spin at 4696 RCF.


There are stickers with maximum speeds. These were set by Katya, circa 2011.  It is unknown what the motivation was for keeping the rotors below maximum speed.
=== SS-34 ===
[[image:Lidstrom_sorvoll_speed_instructions.png|thumb|upright=2.0|center|Lidstrom sorvoll speed instructions]]
* Holds 8*50 mL, spins at '''[http://www.geminibv.nl/labware/sorvall-gsa-rotor/sorvall-rotor-gsa-gs3-sa600-se12-sm24-ss34-manual.pdf 20,000 RPM/36,900 RCF]'''
** BUT Katya (circa 2010) recommends using it at <= 18,000 RPM
**Tubes: [[image:Lidstrom_SS-34 tubes.png|thumb|upright=0.3|center|SS-34 tubes]]  
***Can't be autoclaved. 
**** Actually, Amanda might have autoclaved them before. 
***Can't write on body; can only write on cap.  Sandy says stickers don't stick.


[[image:SS-34 rotor speeds and k factors.png|thumb|upright=2.0|center|SS-34 rotor speeds and k factors]]
[[image:SS-34 rotor speeds and k factors.png|thumb|upright=2.0|center|SS-34 rotor speeds and k factors]]
=== GS-3 ===
* [http://www.thermoscientific.fr/eThermo/CMA/PDFs/Various/File_582.pdf GS-3].
** BUT Katya (circa 2010) recommends using it at <= 7,500
** For bigger bottles.  (0.5L)
** Holds eppendorf (1.5mL) tubes
* Spins at '''[http://www.geminibv.nl/labware/sorvall-gsa-rotor/sorvall-rotor-gsa-gs3-sa600-se12-sm24-ss34-manual.pdf 9,000 RPM/13,700 RCF]'''


[[image:GS-3 rotor speeds and k factors.png|thumb|upright=2.0|center|GS-34 rotor speeds and k factors]]
[[image:GS-3 rotor speeds and k factors.png|thumb|upright=2.0|center|GS-34 rotor speeds and k factors]]


== Explanation of RCF ==
[[image:RCF_explanation.png|thumb|upright=2.0|center|RCF explanation]]
[[image:RCF_explanation.png|thumb|upright=2.0|center|RCF explanation]]
* For reference, the benchtop 50mL centrifuges (e.g. [https://static.thermoscientific.com/images/D21714~.pdf ST-15R] with a [https://extranet.fisher.co.uk/webfiles/fr/Pjointes/Documentation/SOR007_EN%20SORVALL%20ST16&ST40%20CENTRIFUGE.pdf TX-400] swinging bucket rotor) we have spin at 4696 RCF.
*There are stickers with maximum speeds.  These were set by Katya, a Lidstrom Lab manager, circa 2011.  It is unknown what the motivation was for keeping the rotors below maximum speed.
[[image:Lidstrom_sorvoll_speed_instructions.png|thumb|upright=2.0|center|Lidstrom sorvoll speed instructions]]


== Instructions ==
== Instructions ==

Revision as of 17:57, 21 January 2015

Back to Protocols


Background

Lidstrom Sorvoll centrifuge

We have a Sorvoll RC-5B centrifuge that is useful for spinning large volumes in 0.5L bottles or medium-sized (a few mL) volumes very quickly

Rotors

Sorvall rotors


SS-34

  • Holds 8*50 mL, spins at 20,000 RPM/36,900 RCF
    • BUT Katya (circa 2010) recommends using it at <= 18,000 RPM
    • Tubes:
      SS-34 tubes
      • Can't be autoclaved.
        • Actually, Amanda might have autoclaved them before.
      • Can't write on body; can only write on cap. Sandy says stickers don't stick.
SS-34 rotor speeds and k factors

GS-3

  • GS-3.
    • BUT Katya (circa 2010) recommends using it at <= 7,500
    • For bigger bottles. (0.5L)
    • Holds eppendorf (1.5mL) tubes
  • Spins at 9,000 RPM/13,700 RCF
GS-34 rotor speeds and k factors

Explanation of RCF

RCF explanation
  • For reference, the benchtop 50mL centrifuges (e.g. ST-15R with a TX-400 swinging bucket rotor) we have spin at 4696 RCF.
  • There are stickers with maximum speeds. These were set by Katya, a Lidstrom Lab manager, circa 2011. It is unknown what the motivation was for keeping the rotors below maximum speed.
Lidstrom sorvoll speed instructions

Instructions

Example protocol for AM1:

Janet's SIP3-4 centrifugation:

  • Chill to 4oC with SS-34 rotor inside
  • Fill correct tubes
  • Spin for ___ min at ___ RCF at ___oC

Centrifugation Math & Basics

  • Sigma summary:
    • sedimentation rate is proportional to RFC, and proportional to the square of the particle diameter (cell diameter).
    • Stoke's equation
    • centrifugation speeds for applications