Lidstrom:Sorvall Centrifuge: Difference between revisions

From OpenWetWare
Jump to navigationJump to search
Line 8: Line 8:
[http://www.geminibv.nl/labware/sorvall-gsa-rotor/sorvall-rotor-gsa-gs3-sa600-se12-sm24-ss34-manual.pdf Sorvall rotors]
[http://www.geminibv.nl/labware/sorvall-gsa-rotor/sorvall-rotor-gsa-gs3-sa600-se12-sm24-ss34-manual.pdf Sorvall rotors]


Rotors we have:
 
* SS-34Holds 8*50 mL, spins at 20,000 RPM/36,900 RCF
=== SS-34 ===
* Holds 8*50 mL, spins at 20,000 RPM/36,900 RCF
**Tubes: [[image:Lidstrom_SS-34 tubes.png|thumb|upright=0.3|center|SS-34 tubes]]  
**Tubes: [[image:Lidstrom_SS-34 tubes.png|thumb|upright=0.3|center|SS-34 tubes]]  
***Can't be autoclaved.   
***Can't be autoclaved.   
Line 25: Line 26:
[[image:SS-34 rotor speeds and k factors.png|thumb|upright=2.0|center|SS-34 rotor speeds and k factors]]
[[image:SS-34 rotor speeds and k factors.png|thumb|upright=2.0|center|SS-34 rotor speeds and k factors]]


=== GS-3 ===
[[image:GS-3 rotor speeds and k factors.png|thumb|upright=2.0|center|GS-34 rotor speeds and k factors]]
[[image:GS-3 rotor speeds and k factors.png|thumb|upright=2.0|center|GS-34 rotor speeds and k factors]]
[[image:RCF_explanation.png|thumb|upright=2.0|center|RCF explanation]]


== Instructions ==
== Instructions ==

Revision as of 17:45, 21 January 2015

Back to Protocols


Picture

Lidstrom Sorvoll centrifuge

Rotors

Sorvall rotors


SS-34

  • Holds 8*50 mL, spins at 20,000 RPM/36,900 RCF
    • Tubes:
      SS-34 tubes
      • Can't be autoclaved.
        • Actually, Amanda might have autoclaved them before.
      • Can't write on body; can only write on cap. Sandy says stickers don't stick.
  • GS-3.
    • For bigger bottles.
  • ???.
    • Holds eppendorf (1.5mL) tubes
  • For reference, the benchtop 50mL centrifuges (e.g. ST-15R with a TX-400 swinging bucket rotor) we have spin at 4696 RCF.
  • There are stickers with maximum speeds. These were set by Katya, a Lidstrom Lab manager, circa 2011. It is unknown what the motivation was for keeping the rotors below maximum speed.
Lidstrom sorvoll speed instructions
SS-34 rotor speeds and k factors


GS-3

GS-34 rotor speeds and k factors

Instructions

Example protocol for AM1:

Janet's SIP3-4 centrifugation:

  • Chill to 4oC with SS-34 rotor inside
  • Fill correct tubes
  • Spin for ___ min at ___ RCF at ___oC

Centrifugation Math & Basics

  • Sigma summary:
    • sedimentation rate is proportional to RFC, and proportional to the square of the particle diameter (cell diameter).
    • Stoke's equation
    • centrifugation speeds for applications