HighPoint/CannonLab:UV-Vis Spectrometry
From OpenWetWare
(Difference between revisions)
(New page: ==UV-Vis Spectrometry== ===Using the Shimadzu UV-Vis Spectrometer=== ====Startup==== # Open the LCD screen on the spectrometer # Turn ON the spectrometer # Wait for >3.5 minutes for the ...) |
Current revision (12:33, 6 February 2012) (view source) |
||
| Line 1: | Line 1: | ||
| + | {{CannonLabMenu}} | ||
| + | |||
==UV-Vis Spectrometry== | ==UV-Vis Spectrometry== | ||
===Using the Shimadzu UV-Vis Spectrometer=== | ===Using the Shimadzu UV-Vis Spectrometer=== | ||
Current revision
Contents |
UV-Vis Spectrometry
Using the Shimadzu UV-Vis Spectrometer
Startup
- Open the LCD screen on the spectrometer
- Turn ON the spectrometer
- Wait for >3.5 minutes for the instrument to initialize
- If initialization stalls, turn OFF the spectrometer and repeat until the spectrometer completes the checklist (2-3 times is typical)
- Press F4 on the spectrometer to turn control over to the PC
- At any time turn ON the computer
- Open the software
- On the menu option "Configure > Utilities" make sure "spectrophotometer ON" is checked
Settings
Wavelength Scans
Kinetic Measurements