Freimoser:Protocols:ProteinQuantification: Difference between revisions

From OpenWetWare
Jump to navigationJump to search
No edit summary
No edit summary
Line 3: Line 3:
===Source===
===Source===
The quantification methods listed below represent those that we find most useful. The descriptions are part of a more detailed list of protein quantification methods [http://dwb.unl.edu/Teacher/NSF/C10/C10Links/www.ruf.rice.edu/~bioslabs/methods/protein/protein.html]<br>
The quantification methods listed below represent those that we find most useful. The descriptions are part of a more detailed list of protein quantification methods [http://dwb.unl.edu/Teacher/NSF/C10/C10Links/www.ruf.rice.edu/~bioslabs/methods/protein/protein.html]<br>
General Reference: Stoscheck, CM. Quantitation of Protein. Methods in Enzymology 182: 50-69 (1990).
General Reference: Stoscheck, CM. Quantitation of Protein. Methods in Enzymology 182: 50-69 (1990). [http://www.sciencedirect.com/science?_ob=ArticleURL&_udi=B7CV2-4B5XNYK-1Y8&_user=791130&_handle=V-WA-A-W-ZZ-MsSAYWA-UUW-U-AAZVDYAEBB-AAZWBZWDBB-WVEVBEBE-ZZ-U&_fmt=full&_coverDate=12%2F31%2F1990&_rdoc=6&_orig=browse&_srch=%23toc%2318066%231990%23998179999%23473653!&_cdi=18066&view=c&_acct=C000043379&_version=1&_urlVersion=0&_userid=791130&md5=2124740cecad6e53b521353d92a0958c]
 


==Absorbance assays==
==Absorbance assays==

Revision as of 08:22, 30 October 2006

<html><style type='text/css'> A:link {

   text-decoration: underline;
   font-size: 10pt;
   font-style: normal;
   font-family: arial, trebuchet ms;
   color: #003399;

}

A:visited {

   text-decoration: underline;
   font-size: 10pt;
   font-style: normal;
   font-family: arial, trebuchet ms;
   color: #003399;

} A:active {

   text-decoration: none;
   background-color: #99ccff;
   color: #660099;

}

A:hover {

   text-decoration: none;
   color: #ff0033;

}

.navbar{

 padding: 0px;
 width: "100%";
 font-family: Arial;
 font-size: 10pt;

} </style></html>

Source

The quantification methods listed below represent those that we find most useful. The descriptions are part of a more detailed list of protein quantification methods [1]
General Reference: Stoscheck, CM. Quantitation of Protein. Methods in Enzymology 182: 50-69 (1990). [2]


Absorbance assays

Absorbance at 280 nm

  • Range: 20 micrograms to 3 mg
  • Volume: Depends on cuvette - volumes range from 200 microliters to 3 ml or greater
  • Accuracy: Fair
  • Convenience: Excellent, if equipment available
  • Major interfering agents: Detergents, nucleic acids, particulates, lipid droplets

Absorbance at 205 nm

  • Range: Roughly 1 to 100 micrograms
  • Volume: Depends on cuvette - volumes range from 200 microliters to 3 ml or greater
  • Accuracy: Fair
  • Convenience: Very good
  • Major interfering agents: Detergents, nucleic acids, particulates, lipid droplets


Colorimetric assays

Modified Lowry

  • Range: 2 to 100 micrograms
  • Volume: 1 ml (scale up for larger cuvettes)
  • Accuracy: Good
  • Convenience: Fair
  • Major interfering agents: Strong acids, ammonium sulfate

Bradford assay

  • Range: 1 to 20 micrograms (micro assay); 20 to 200micrograms (macro assay)
  • Volume: 1 ml (micro); 5.5 ml (macro)
  • Accuracy: Good
  • Convenience: Excellent
  • Major interfering agents: None