BioMicroCenter:Sequencing Quality Control: Difference between revisions
From OpenWetWare
Jump to navigationJump to search
Line 11: | Line 11: | ||
* [[BioMicroCenter:2100BioAnalyzer|Bioanalyzer]] - This method quantifies samples sequentially by their size by molecular sieving and are then detected by fluorescence. This method is only suggested for samples suggested for samples exceeding 10ng/ul and has shown to be less reliable at lower concentrations. Available at the BMC for $10 per sample | * [[BioMicroCenter:2100BioAnalyzer|Bioanalyzer]] - This method quantifies samples sequentially by their size by molecular sieving and are then detected by fluorescence. This method is only suggested for samples suggested for samples exceeding 10ng/ul and has shown to be less reliable at lower concentrations. Available at the BMC for $10 per sample | ||
* Qubit - This method quantifies samples by using fluorescent dyes that are specific for dsDNA, RNA or protein. This method is used in the Fraenkel lab to quantify their samples for sequencing. Not yet tested at the BMC but information about this application can be found at http://openwetware.org/index.php?title=BioMicroCenter:Sequencing_Quality_Control&action=edit&redlink=1 | |||
* PicoGreen - This method makes use of PicoGreen and fluorescent dye that binds specifically to dsDNA. The fluorescence of this dye can by measured a few different ways: the manufacturer states that the Qubit can be used, the B lab uses the photospectrometer | |||
* RT-PCR, SYBERgreen assay - This assay allows for the amount of DNA that will actually bind to the flowcell to be quantified. This is achieved by making use of the flowcell primers in the QPCR assay. Coming soon to the BMC!! | * RT-PCR, SYBERgreen assay - This assay allows for the amount of DNA that will actually bind to the flowcell to be quantified. This is achieved by making use of the flowcell primers in the QPCR assay. Coming soon to the BMC!! | ||
Revision as of 13:50, 12 January 2009
HOME -- | SEQUENCING -- | LIBRARY PREP -- | HIGH-THROUGHPUT -- | COMPUTING -- | OTHER TECHNOLOGY |
Why is QC Important?
It is very important to supply us with a reliable measurement of the concentration and quantity of starting material so that we can prepare the sample for sequencing. If a sample is too concentrated the cluster density will be too high and the GA will not be able to detect them properly. If a sample is too dilute the cluster density will be too low and there will not be enough reads for the pipeline to make base calls. Having reliable QC measurements allows us to optimize the cluster density and the number of reads per lane and maximize the quality of data produced. QC can either be run before sample submission and specified upon testing request or can be run by the BMC. We strongly recommend that you allow us to run QC of samples.
Possible QC Techniques
- Nanodrop - this method allows you to quantify samples without diluting them. It has a detection limit of about 2ng/ul to 3700ng/ul. However, the Nanodrop is not recommended for sequencing samples because this method is not specific to dsDNA and has proven to have great variation at lower concentrations. Use of the BMC Nanodrop is FREE
- Bioanalyzer - This method quantifies samples sequentially by their size by molecular sieving and are then detected by fluorescence. This method is only suggested for samples suggested for samples exceeding 10ng/ul and has shown to be less reliable at lower concentrations. Available at the BMC for $10 per sample
- Qubit - This method quantifies samples by using fluorescent dyes that are specific for dsDNA, RNA or protein. This method is used in the Fraenkel lab to quantify their samples for sequencing. Not yet tested at the BMC but information about this application can be found at http://openwetware.org/index.php?title=BioMicroCenter:Sequencing_Quality_Control&action=edit&redlink=1
- PicoGreen - This method makes use of PicoGreen and fluorescent dye that binds specifically to dsDNA. The fluorescence of this dye can by measured a few different ways: the manufacturer states that the Qubit can be used, the B lab uses the photospectrometer
- RT-PCR, SYBERgreen assay - This assay allows for the amount of DNA that will actually bind to the flowcell to be quantified. This is achieved by making use of the flowcell primers in the QPCR assay. Coming soon to the BMC!!