BioMicroCenter:SPRI-Works: Difference between revisions

From OpenWetWare
Jump to navigationJump to search
Line 9: Line 9:
The SPRI-works system  consists of three main components: the SPRI-TE Nucleic Acid Extractor, a method card that controls liquid handling, and a cartridge containing all the reagents required to prepare a single library.  
The SPRI-works system  consists of three main components: the SPRI-TE Nucleic Acid Extractor, a method card that controls liquid handling, and a cartridge containing all the reagents required to prepare a single library.  


This platform utilizes DNA binding magnetic beads to purify fragmented DNA and perform size selection, replacing the need for column purification and gel-base size selection. There are two size range options available, 100-400bp or 300-600bp. There is also the choice to perform no size selection
This platform utilizes DNA binding magnetic beads to purify fragmented DNA and perform size selection, replacing the need for column purification and gel-base size selection. There are two size range options available, 100-400bp or 300-600bp. There is also the choice to perform no size selection. For more information on the SPRI-Works system please visit www.spriworks.com


==Sample Submission==
==Sample Submission==

Revision as of 12:52, 2 August 2010

HOME -- SEQUENCING -- LIBRARY PREP -- HIGH-THROUGHPUT -- COMPUTING -- OTHER TECHNOLOGY

The BMC is currently testing the Beckman SPRIworks library preparation system. This platform is a fully automated library preparation system for Illumina sequencing and supports all Illumina applications except for smallRNA-Seq. It performs all of the traditional Illumina protocol steps listed on the Illumina Library Preparation page, except for Enrichment, can prepare up to ten libraries for Illumina sequencing in 5 hours.

How it Works

The SPRI-works system consists of three main components: the SPRI-TE Nucleic Acid Extractor, a method card that controls liquid handling, and a cartridge containing all the reagents required to prepare a single library.

This platform utilizes DNA binding magnetic beads to purify fragmented DNA and perform size selection, replacing the need for column purification and gel-base size selection. There are two size range options available, 100-400bp or 300-600bp. There is also the choice to perform no size selection. For more information on the SPRI-Works system please visit www.spriworks.com

Sample Submission

All samples submitted for direct SPRI-Works library preparation must be submitted with user supplied adapter mix. Please note that samples will need to be submitted separately for Illumina sequencing.


Sample Submission Forms