BioMicroCenter:SPRI-Works: Difference between revisions

From OpenWetWare
Jump to navigationJump to search
No edit summary
Line 16: Line 16:
==Sample Submission==
==Sample Submission==


Sample concentrations for the SPRI-TE robot are highly variable, we have successfully run samples as low as 0.3ng/uL. It is suggested that you submit as much volume as possible, as the first step of the protocol is to bring all sample volumes up to 100uLs in water.  
Input sample concentrations on the SPRI-te are variable (we have successfully run samples as low as 0.3ng/uL). We suggest that you submit as much volume as possible, as the first step of the protocol is to bring all sample volumes up to 100uL in water.  


All samples submitted for '''BMC Fragmented DNA sample prep (SPRI-only)''' must be submitted with '''user supplied''' adapter mix. Adapter mixes should be pre-diluted and should have enough volume for 10uL of adapter per sample. For recommended dilutions for various sample concentrations please see the chart below:
All samples submitted for '''BMC Fragmented DNA sample prep (SPRI-only)''' must be submitted with '''user supplied''' adapter mix. Adapter mixes should be pre-diluted and should have enough volume for 10uL of adapter per sample. For recommended dilutions for various sample concentrations, please see the chart below:


[[Image:Spri-adaptrecomend.jpg|center|500px]]
[[Image:Spri-adaptrecomend.jpg|center|500px]]




Please note that samples will need to be resubmitted separately for Illumina sequencing.
Please note that SPRI-only samples will need to be re-submitted separately if you wish to sequence with us later on.


[[BioMicroCenter:Forms#Downloads|Sample Submission Forms]]
[[BioMicroCenter:Forms#Downloads|Sample Submission Forms]]

Revision as of 12:59, 28 September 2011

HOME -- SEQUENCING -- LIBRARY PREP -- HIGH-THROUGHPUT -- COMPUTING -- OTHER TECHNOLOGY

The SPRI-te Nucleic Acid Extractor is a fully-automated library preparation system for Illumina sequencing and supports all Illumina applications except for smallRNA-Seq. It performs all of the traditional Illumina protocol steps listed on the Illumina Library Preparation page with the exception of PCR enrichment. A total of 10 libraries may be prepared simultaneously in 5 hours.

How it Works

The SPRI-works system consists of three main components: the SPRI-TE Nucleic Acid Extractor, a method card that contains the liquid handling program, and a cartridge containing all of the reagents required to prepare a single library.

The platform utilizes AMPureXP DNA-binding magnetic beads to purify fragmented DNA and perform size selection, eliminating the need for column purification and gel-based size selection. There are three size range options available, 200-400bp, 300-600bp, or no size selection.
After the adapter ligation and size selection are complete, samples are ready for PCR enrichment and clean-up to complete the library prep process.

For more information on the SPRI-Works system please visit www.spriworks.com.

Sample Submission

Input sample concentrations on the SPRI-te are variable (we have successfully run samples as low as 0.3ng/uL). We suggest that you submit as much volume as possible, as the first step of the protocol is to bring all sample volumes up to 100uL in water.

All samples submitted for BMC Fragmented DNA sample prep (SPRI-only) must be submitted with user supplied adapter mix. Adapter mixes should be pre-diluted and should have enough volume for 10uL of adapter per sample. For recommended dilutions for various sample concentrations, please see the chart below:


Please note that SPRI-only samples will need to be re-submitted separately if you wish to sequence with us later on.

Sample Submission Forms