BioMicroCenter:Advanced analytical Fragment analyzer: Difference between revisions

From OpenWetWare
Jump to navigationJump to search
Line 48: Line 48:
  |-
  |-
  !Prep/Loading Time
  !Prep/Loading Time
  |~5' for 12 samples
  |<5' for 12 samples
  |>30' for 10 samples
  |>30' for 10 samples
  |-
  |-

Revision as of 06:40, 6 August 2013

Purchased in July of 2012, the Advanced analytical Fragment analyzer is a highly efficient and accurate fluorescence-based electrophoresis instrument used to size and quantify both high sensitivity DNA and Nano RNA (conc. >5ng/ul). It is faster and more cost effective than the BioAnalyzer2100, and can handle unsupervised runs of up to 96 samples. With a failure rate of less than one percent, the machine’s efficiency allows the BMC to analyze most samples within 24 hours of submission. A run of 12 samples takes 60 minutes and 2 μL of sample with a cost of $8 per sample.

DNA

The Advanced analytical analyzes DNA using capillary electrophoresis. A tiny amount of sample is injected through a capillary with gel, and across a fluorescence detection window using a voltage gradient. DNA intercalating dye within the gel makes the DNA detectable by a camera lens, which is able to read the fluorescence levels of the fragments through the window, and accurately give the size and concentrations of the sample. For a more in-depth description see the link below:

Fragment Analyzer - Parallel Capillary Electrophoresis (Advanced Analytical)

The Advanced analytical processes DNA fragments of 50-5000bp at a wide range of concentrations:

5pg/μL-1000pg/μL (Fragments)

50pg/μL-5000pg/μL (Smear)


NGS kit: 50bp-5000bp, 5pg/ul-1000pg/μL

RNA

The Advanced analytical analyzes RNA using capillary electrophoresis. A tiny amount of sample is injected through a capillary with gel, and across a fluorescence detection window using a voltage gradient. RNA intercalating dye within the gel makes the RNA detectable by a camera lens, which is able to read the fluorescence levels of the fragments through the window, and accurately give the size and concentrations of the sample.

The Advanced analytical provides an RNA quality number (RQN) from 1-10 on samples of Nano RNA. It is capable of analyzing samples at concentrations of 5ng/ul-500ng/ul.

Standard sensitivity RNA kit: 5ng/μL-500ng/μL

Problem Identification

The Advanced Analytical is capable of identifying problems such as hyperamplification and “primer-dimer”.

Platform Comparison with BioAnalyzer 2100

Advanced Analytical BioAnalyzer
Volume per sample 2 μL 1 μL
Time per run 60' for 12 samples (up to 96 samples) 45' for 10 samples
Prep/Loading Time <5' for 12 samples >30' for 10 samples
Failure Rate <1% 15-30%
Minimum Conc. ~.06nM (DNA) ~.02nM (DNA)
Cost Per Sample $8 $15
Assays High Sens. DNA & Nano RNA Many