Beauchamp:NewEyeTrackSetup: Difference between revisions
From OpenWetWare
Jump to navigationJump to search
Nathan Doyle (talk | contribs) |
|||
Line 18: | Line 18: | ||
Display PC needs to have IP address manually set to 100.1.1.2, and the subnet mask to 255.255.255.0 | Display PC needs to have IP address manually set to 100.1.1.2, and the subnet mask to 255.255.255.0 | ||
- use any web browser to access files on Host Laptop by going to 100.1.1.1 in address bar | - use any web browser to access files on Host Laptop by going to 100.1.1.1 in address bar | ||
To run n experiment, have the host PC (the laptop) on and have the EyeLink camera running before starting the experiment program, otherwise the program will time out. | |||
Setting up for new monitor is simple: go to the settings on the Host PC, click the screen setup icon, and enter the measurements it asks for. |
Revision as of 12:34, 13 January 2014
Notes on the SR EyeLink Eye Tracker
The Beauchamp Lab has two SR Eyelink systems: an Eyelink (for MRI scanner use) and an Eyelink Plus (set up in the lab).
Here are notes on setting up the EyeLink Plus:
- connections:
- laptop to camera via ethernet - laptop to display PC via ethernet (USB ethernet port) - illuminator to camera
Ethernet connections must be in place before Host Laptop is turned on.
Display PC needs to have IP address manually set to 100.1.1.2, and the subnet mask to 255.255.255.0
- use any web browser to access files on Host Laptop by going to 100.1.1.1 in address bar
To run n experiment, have the host PC (the laptop) on and have the EyeLink camera running before starting the experiment program, otherwise the program will time out.
Setting up for new monitor is simple: go to the settings on the Host PC, click the screen setup icon, and enter the measurements it asks for.