BME100 s2014:T Group10 L5: Difference between revisions
(39 intermediate revisions by 2 users not shown) | |||
Line 29: | Line 29: | ||
'''SYBR Green Dye'''<br> | '''SYBR Green Dye'''<br> | ||
SYBR Green Dye is a dye that stains nucleic acids by binding to DNA or RNA. The dye is used to visualize DNA through staining agarose gels, | SYBR Green Dye is a dye that stains nucleic acids by binding to DNA or RNA. The dye is used to visualize DNA through staining agarose gels, | ||
biochip applications, fluorescence imaging techniques, real-time PCR, and more. SYBR dye fluoresces when combined with dsDNA (Zipper Hubert Brunner 1). | biochip applications, fluorescence imaging techniques, real-time PCR, and more. SYBR dye fluoresces when combined with dsDNA (Zipper Hubert Brunner 1). However, unbound SYBR Green is unreactive to light. This feature allows scientists to determine the concentration of a target DNA sample by measuring the intensity of the fluorescent light being emitted by the sample. The higher the intensity of the light being emitted, the more concentrated the sample is with the target DNA. SYBR Green, as its name implies, fluoresces with a green color. Accordingly, it is important to use a different wavelength of light to excite the sample in order to prevent distorting the results. There are several types of fluorescent dyes similar to the SYBR Green used in this lab, but SYBR Green was chosen for its environmental friendliness and high level of light emission intensity to allow for easier detection of binding to the target DNA.<br> | ||
'''Single-Drop Fluorimeter'''<br> | '''Single-Drop Fluorimeter'''<br> | ||
A fluorimeter is a device used to measure and identify fluorescence in a medium. Fluorimeters measure certain wavelengths of light and have two light detectors, | A fluorimeter is a device used to measure and identify fluorescence in a medium. Fluorimeters measure certain wavelengths of light and have two light detectors, one of which detects absorbency while the other detects fluorescence emission (So Dong 1). Single-Drop Fluorimeters do this with a single drop of a sample that is placed on a hydrophobic slide in the path of a beam of single wavelength light. For example, in our setup we used blue LED light to allow for light absorption and prevent contaminated intensity data that would occur if we used a green LED since the SYBR Green Dye fluoresces in the green range. A light detector or camera placed at a uniform distance from this sample is then used to capture the intensity of the fluorescence being emitted from the sample. Cameras are acceptable for this laboratory set-up even though they are unable to measure light intensity because it is assumed that light intensity is proportional to brightness; the brighter an object is in a photo, the higher the light intensity. | ||
one of which detects absorbency while the other detects fluorescence emission (So Dong 1). | |||
[[Image: BME100_S14_G10Th_Naked_Fluorimeter.JPG |300px|Single-Drop Fluorimeter setup without covering.]]<br> | |||
A | (Figure: Our Single-Drop Fluorimeter setup. A sample is placed in the path of the LED light on the white hydrophobic slide and its light intensity is captured by taking a picture with the iPhone. Image by Nathan Kirkpatrick)<br><br> | ||
[[Image: BME100_S14_G10Th_Covered_Fluorimeter.JPG |300px|Single-Drop Fluorimeter setup with covering.]]<br> | |||
(Figure: Our Single-Drop Fluorimeter covered by a box to prevent ambient light interference. A timer is used to take the picture of the sample once the box is closed. Image by Nathan Kirkpatrick)<br><br> | |||
'''How the Fluorescence Technique Works'''<br> | |||
In this experiment, the concentration of DNA was determined by measuring the amount of fluorescence in a medium that had been dyed with SYBR Green Dye. Since light intensity is assumed to be proportional to brightness, images taken with a camera using the Single-Drop Fluorimeter setup described above can be used to determine the intensity of the SYBR Green Dye and, accordingly, the concentration of target DNA in the sample. After a single-drop fluorimeter was used to measure the fluorescence and an image was taken, ImageJ software was used to determine the concentration of DNA. We use the ImageJ software to split the color channels and work with the green channel, thereby eliminating intensity contamination from the blue LED. Then we draw an oval around the drop and use the software to calculate the RawIntDen of the drop that is the intensity reading of the part of the image within the oval (the drop). Since the pictures were taken at a uniform distance from the drop and the total volume of each drop is the same, we can assume that the area of the drop is constant and are thusly able to compare the intensity values directly without any conversion. For the calibration, we also took a background sample oval of the same area in ImageJ to get a baseline intensity for that image that was subtracted from the intensity value of the drop to get more accurate data. The intensity of the drops of patient samples are then compared to positive and negative controls to determine if they contain the target DNA sequences. For the PCR Data, we conducted a quick calibration of three SYBR Green I and DNA sample concentrations then plotted their intensities on a graph. The line of best fit for this graph can then be used to determine the concentration of target DNA sequences in for our PCR samples with unknown amounts of the target DNA sequences. | |||
<br> | |||
==Procedure== | ==Procedure== | ||
'''Smart Phone Camera Settings'''<br> | '''Smart Phone Camera Settings'''<br> | ||
* Type of Smartphone: Apple iPhone5S | |||
* Type of Smartphone: | ** Flash: no flash | ||
** Flash: | ** ISO setting: 2500 | ||
** ISO setting: | ** White Balance: Auto | ||
** White Balance: | ** Exposure: Highest Setting | ||
** Exposure: | ** Saturation: N/A | ||
** Saturation: | ** Contrast: N/A | ||
** Contrast: | |||
'''Calibration'''<br> | '''Calibration'''<br> | ||
* Distance between the smart phone cradle and drop = 8cm | |||
# Place a slide with a hydrophobic side in the fluorimeter tray with the hydrophobic side up. | |||
# Turn on the blue LED light and make sure it is aligned with the sample drop. | |||
# Prepare the camera by ensuring the drop is in focus, centered in the frame, the flash is off, and the ISO is locked at 2500 (we were unable to control the other settings listed in the laboratory workbook. | |||
# Verify that the distance from the drop to the camera is 8 cm. | |||
# Place 80 μL drop in between two holes in the hydrophobic slide to prevent it from moving. | |||
# Add 80 μL of the calf thymus solution or water blank on top of the SYBR Green drop. | |||
# Make sure it is lined up with the blue LED light. | |||
# Double check that the iPhone camera is 8 cm from the drop. | |||
# Set the timer of the iPhone camera to 5 seconds. | |||
# Place the black-out box over the whole set-up carefully. | |||
# Close the cover. | |||
# Check to make sure picture is in focus and valid. | |||
# Check the iPhone-Drop distance again. | |||
# Take two more pictures. | |||
# Remove the drop with a pipette. | |||
# Repeat these step for all calibration concentrations using different locations on the hydrophobic slide for each new concentration.<br> | |||
[[Image: BME100_S14_G10Th_Fluorimeter_Above.JPG |300px|Aerial view of Single-Drop Fluorimeter Set-up. the distance from the iPhone's camera to the drop is 8 cm.]]<br> | |||
(Figure: Aerial view of Single-Drop Fluorimeter Set-up. the distance from the iPhone's camera to the drop is 8 cm.)<br><br> | |||
'''Solutions Used for Calibration''' | '''Solutions Used for Calibration'''<br> | ||
{| {{table}} | {| {{table}} | ||
| align="center" style="background:#f0f0f0;"|'''Initial Concentration of 2X Calf Thymus DNA Solution (μg/mL)''' | |||
| align="center" style="background:#f0f0f0;"|'''Volume of the 2X DNA Solution (μL)''' | |||
| align="center" style="background:#f0f0f0;"|'''Volume of the SYBR Green I Dye Solution (μL)''' | |||
| align="center" style="background:#f0f0f0;"|'''Final DNA Concentration in SYBR Green I Solution (μg/mL)''' | |||
|- | |||
| 5||80||80||2.5 | |||
|- | |||
| 2||80||80||1 | |||
|- | |- | ||
| | | 1||80||80||0.5 | ||
|- | |- | ||
| | | 0.5||80||80||0.25 | ||
|- | |- | ||
| | | 0.25||80||80||0.125 | ||
|- | |||
| 0||80||80||0 | |||
|} | |} | ||
'''Placing Samples onto the Fluorimeter''' | '''Placing Samples onto the Fluorimeter''' | ||
# | # Make sure you are wearing proper personal protective equipment. | ||
# | # Make sure the hydrophobic slide has the hydrophobic side face up. | ||
# | # Make sure slide is lined up with the LED. | ||
# | # Make sure micropipette is set to 80 μL. | ||
# Use good micropipette practices to add 80 μL of SYBR Green I to in between two holes in the middle column of the hydrophobic slide. | |||
# Use good micropipette practices to add 80 μL of the calibration solution on top of this drop. | |||
# Follow steps listed above to get pictures. | |||
# Use pipette to remove the dot. | |||
# Put used solution mix into liquid biohazard waste container. | |||
# Move slide to the next set of holes. | |||
# Repeat for each calibration concentration. | |||
# When finished place the slide in the biohazard sharps container. | |||
# Dispose of pipette, micropipette tips, gloves and other used materials in the biohazard container. | |||
<br> | <br> | ||
==Data Analysis== | ==Data Analysis== | ||
'''Representative Images of Samples''' | '''Representative Images of Samples'''<br><br> | ||
[[Image:BME100_S14_G10Th_NC_ImageJ.png|300px|ImageJ Green-Channel isolated Negative Control image with circle around droplet.]]<br> | |||
(Figure: ImageJ Green-Channel isolated Negative Control image with circle around droplet)<br><br> | |||
[[Image:BME100_S14_G10Th_PC_ImageJ.png|300px|ImageJ Green-Channel isolated Positive Control image with circle around droplet.]]<br> | |||
(Figure: ImageJ Green-Channel isolated Positive Control image with circle around droplet)<br><br> | |||
'''Image J Values for All Samples''' | '''Image J Values for All Samples''' <br><br> | ||
'''Calibration Data'''<br> | |||
{| {{table}} | |||
| align="center" style="background:#f0f0f0;"|'''Final DNA Concentration in SYBR Green I Solution (μg/mL)''' | |||
| align="center" style="background:#f0f0f0;"|'''Area''' | |||
| align="center" style="background:#f0f0f0;"|'''Mean Pixel Value''' | |||
| align="center" style="background:#f0f0f0;"|'''RAWINTDEN of the Drop''' | |||
| align="center" style="background:#f0f0f0;"|'''RAWINTDEN of the Background''' | |||
| align="center" style="background:#f0f0f0;"|'''Subtracted RAWINTDEN''' | |||
|- | |||
| 0||35438||29.469||1044333||5349||1038984 | |||
|- | |||
| 0||35508||28.895||1026007||4082||1021925 | |||
|- | |||
| 0||38040||27.429||1043189||2237||1040952 | |||
|- | |||
| 0||33727||28.897||974620||8357||966263 | |||
|- | |||
| 0.25||34864||56.034||1953555||10309||1943246 | |||
|- | |||
| 0.25||36644||47.857||1753686||11015||1742671 | |||
|- | |||
| 0.25||32122||45.717||1468518||6105||1462413 | |||
|- | |||
| 0.25||31920||46.667||1489608||6226||1483382 | |||
|- | |||
| 0.5||36018||66.312||2388411||6733||2381678 | |||
|- | |||
| 0.5||35838||66.157||2370925||7255||2363670 | |||
|- | |||
| 0.5||34096||70.996||4240685||10984||2409701 | |||
|- | |||
| 0.5||33313||64.808||2158943||7057||2151886 | |||
|- | |||
| 1||32710||100.82||3297817||7473||3290344 | |||
|- | |||
| 1||32380||101.547||3288080||7660||3280420 | |||
|- | |||
| 1||36709||87.955||3228755||7600||3221155 | |||
|- | |||
| 1||37464||87.787||3288829||8455||3280374 | |||
|- | |||
| 2||36244||112.346||4071880||11409||4060471 | |||
|- | |||
| 2||33123||105.502||3494534||30121||3464413 | |||
|- | |||
| 2||30036||105.369||3164875||16870||3148005 | |||
|- | |||
| 2||27950||115.15||3246380||5312||3241078 | |||
|- | |||
| 5||31172||129.814||4046570||11325||4035245 | |||
|- | |||
| 5||30780||130.063||4003335||11554||3991781 | |||
|- | |||
| 5||33575||128.746||4322651||4541||4318110 | |||
|- | |||
| 5||33727||129.1||4356861||373||4356132 | |||
|}<br><br> | |||
[[Image: BME100_S14_G10Th_CalibrationPlot.png|600px|Calibration Data Plot]]<br> | |||
(Figure: Plot of the calibration data with Calf Thymus DNA concentration on the x-axis and INTDENS Readings after background correction on the y-axis with trend line.)<br><br> | |||
'''PCR Product Sample Measurements''' | |||
{| {{table}} | |||
| align="center" style="background:#f0f0f0;"|'''PCR Product Tube Label''' | |||
| align="center" style="background:#f0f0f0;"|'''Area (pixel^2)''' | |||
| align="center" style="background:#f0f0f0;"|'''Mean Pixel Value''' | |||
| align="center" style="background:#f0f0f0;"|'''RawIntDen of the Drop''' | |||
|- | |||
| PC 1||30528||123.781||3778777 | |||
|- | |||
| PC 2||28794||124.966||3598262 | |||
|- | |||
| PC 3||28349||126.204||3577769 | |||
|- | |||
| NC 1||26035||49.289||1283252 | |||
|- | |||
| NC 2||25986||38.364||1002239 | |||
|- | |||
| NC 3||24081||34.513||831118 | |||
|- | |||
| 1R1 1 ||26780||40.389||1081609 | |||
|- | |||
| 1R1 2||28678||38.826||1113459 | |||
|- | |||
| 1R1 3||27348||39.13||1070129 | |||
|- | |||
| 1R2 1||27328||43.119||1178362 | |||
|- | |||
| 1R2 2||29178||45.662||1332320 | |||
|- | |||
| 1R2 3||26307||47.495||1249444 | |||
|- | |||
| 1R3 1||22214||44.459||987619 | |||
|- | |||
| 1R3 2||26264||32.101||843096 | |||
|- | |||
| 1R3 3 ||25368||33.248||843428 | |||
|- | |||
| 2R1 1||22296||126.872||2828741 | |||
|- | |||
| 2R1 2||25085||130.523||3274161 | |||
|- | |||
| 2R1 3||24628||130.723||3219456 | |||
|- | |||
| 2R2 1||26100||129.338||3375712 | |||
|- | |||
| 2R2 2||26272||128.394||3373160 | |||
|- | |||
| 2R2 3||23711||129.676||3074751 | |||
|- | |||
| 2R3 1||23900||128.988||3082802 | |||
|- | |||
| 2R3 2||23824||131.695||3137491 | |||
|- | |||
| 2R3 3||27428||131.43||3604871 | |||
|} | |||
<br> | |||
The table below shows the concentrations of the PCR Products. We used the line of best fit from our quick calibration to determine these values. The PCR Product Concentrations was determined by subtracting the Average INTDENS value of the drop by the intercept of the best fit line (2,000,000 INTDENS) and dividing by the slope of the best fit line (1,000,000 INTDENS/concentration). The concentrations then need to be adjusted for the total dilution that took place in the experiment. When 100 μL of the PCR reaction was added to 500 μL of buffer, the PRC reaction wad diluted by 1/6. Additionally, when 80 μL of the diluted PCR reaction was added to 80 μL of SYBR Green I solution, the PCR reaction was further diluted by 1/2. In total, the PCR reaction was diluted by 1/12. Accordingly, the PCR product concentrations need to be adjusted by multiplying by the denominator of the total dilution (12). These values are reflected in the fourth column of this table.<br><br> | |||
{| {{table}} | |||
| align="center" style="background:#f0f0f0;"|'''PCR Product Tube Label''' | |||
| align="center" style="background:#f0f0f0;"|'''Average INTDENS Value of the Drop''' | |||
| align="center" style="background:#f0f0f0;"|'''Average PCR Product Concentration μg/mL''' | |||
| align="center" style="background:#f0f0f0;"|'''Corrected Average PCR Product Concentration, μg/mL''' | |||
|- | |||
| PC||3651602.667||1.651602667||19.819232 | |||
|- | |||
| NC||1038869.667||-0.961130333||-11.533564 | |||
|- | |||
| 1R1||1088399||-0.911601||-10.939212 | |||
|- | |||
| 1R2||1253375.333||-0.746624667||-8.959496 | |||
|- | |||
| 1R3||891381||-1.108619||-13.303428 | |||
|- | |||
| 2R1||3107452.667||1.107452667||13.289432 | |||
|- | |||
| 2R2||3274541||1.274541||15.294492 | |||
|- | |||
| 2R3||3275054.667||1.275054667||15.300656 | |||
|} | |||
<br> | |||
For the sake of clarity, the samples are listed below with their corresponding Patient ID Number and corrected PCR Product Concentration. The Positive and Negative controls are also included.<br> | |||
{| {{table}} | |||
| align="center" style="background:#f0f0f0;"|'''Sample''' | |||
| align="center" style="background:#f0f0f0;"|'''Corrected PCR Product Concentration (μg/ml)''' | |||
|- | |||
| Postive Control||19.819232 | |||
|- | |||
| Negative Control||-11.533564 | |||
|- | |||
| Patent: 21489 Replicate 1||-10.939212 | |||
|- | |||
| Patent: 21489 Replicate 2||-8.959496 | |||
|- | |||
| Patent: 21489 Replicate 3||-13.303428 | |||
|- | |||
| Patent: 90434 Replicate 1||13.289432 | |||
|- | |||
| Patent: 90434 Replicate 2||15.294492 | |||
|- | |||
| Patent: 90434 Replicate 3||15.300656 | |||
|}<br><br> | |||
'''Fitting a Straight Line'''<br> | '''Fitting a Straight Line'''<br> | ||
[[Image: BME100_S14_G10Th_PCR_DataGraph.png|600px|PCR Data Plot]]<br> | |||
(Figure: PCR Product Sample Quick Calibration measurements with a line of best fit.) | |||
<br> | <br> | ||
< | <b>PCR Results Summary</b><br /> | ||
Instructor's summary: You completed 8 PCR reactions in a previous lab. You used the SYBR Green I staining and imaging technique to measure the amount of amplified DNA in each PCR reaction. You used a standard curve (based on known concentrations of calf thymus DNA) to convert INTDEN values into DNA concentration. Your positive control and negative control samples should be used as <b>threshold</b> values for determining whether an unknown (patient) sample is truly positive or negative. | |||
<ul><li> Your positive control PCR result was 19.819232 μg/mL | |||
</li><li> Your negative control PCR result was -11.533564 μg/mL | |||
</li></ul> | |||
<br /> | |||
Write-in each patient ID and give both a qualitative (what the images looked like) and a quantitative description (μg/mL) of what you observed | |||
<ul><li> Patient 21489 : The samples containing this patient's PCR-duplicated DNA with SYBR Green I did not appear to exhibit SYBR Green I-affiliated fluorescence. In other words, the droplet did not glow at all. The Average Corrected PCR Product Concentration of -11.06737867 μg/mL for all three replicates of this patient supports this visual observation. | |||
</li><li> Patient 90434 :The samples containing this patient's PCR-duplicated DNA with SYBR Green I did appear to exhibit SYBR Green I-affiliated fluorescence. In other words, the droplet was visually glowing. The Average Corrected PCR Product Concentration of 14.62819333 μg/mL for all three replicates of this patient supports this visual observation. | |||
</li></ul> | |||
<p>Compare each patient's results to the positive control value and the negative control value. Draw a final conclusion for each patient (positive or negative) and explain why you made that conclusion. | |||
</p> | |||
<ul><li> Patient 21489 : The Average Corrected PCR Product Concentration of -11.06737867 μg/mL for all three replicates of this patient is very close to the negative control threshold value of -11.533564 μg/mL. From this, we can conclude that Patient 21489 does not contain the DNA sequence in question. We came to this conclusion because the SYBR Green I fluoresced as much in this patient's samples as it did in the negative control where no target DNA sequences were present. | |||
</li><li> Patient 90434 : The Average Corrected PCR Product Concentration of 14.62819333 μg/mL for all three replicates of this patient is very close to the positive control threshold value of 19.819232 μg/mL. From this, we can conclude that Patient 21489 does contain the DNA sequence in question. We came to this conclusion because the SYBR Green I fluoresced as much in this patient's samples as it did in the positive control where a significant amount of target DNA sequences were present. Unfortunately, Patient 90434 has the condition associated with this allele. | |||
</li></ul> | |||
==Works Cited== | |||
Zipper, Hubert, Herwig Brunner, Jurgen Bernhagen, and Frank Vitzthum. "Investigations on DNA Intercalation and Surface Binding by SYBR Green I, Its Structure Determination and Methodological Implications." Ncbi.nih.gov. US National Library of Medicine, 12 July 2004. Web. 3 Apr. 2014.<br><br> | |||
So, Peter TC, and Chen Y. Dong. "Fluorescence Spectrophotometry." Scribd. Macmillan Publishers Ltd, Nature Publishing Group, 16 Jan. 2009. Web. 03 Apr. 2014. | |||
<!-- Do not edit below this line --> | <!-- Do not edit below this line --> | ||
|} | |} |
Latest revision as of 09:53, 17 April 2014
BME 100 Spring 2014 | Home People Lab Write-Up 1 | Lab Write-Up 2 | Lab Write-Up 3 Lab Write-Up 4 | Lab Write-Up 5 | Lab Write-Up 6 Course Logistics For Instructors Photos Wiki Editing Help | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
OUR TEAMLAB 5 WRITE-UPBackground InformationSYBR Green Dye SYBR Green Dye is a dye that stains nucleic acids by binding to DNA or RNA. The dye is used to visualize DNA through staining agarose gels,
biochip applications, fluorescence imaging techniques, real-time PCR, and more. SYBR dye fluoresces when combined with dsDNA (Zipper Hubert Brunner 1). However, unbound SYBR Green is unreactive to light. This feature allows scientists to determine the concentration of a target DNA sample by measuring the intensity of the fluorescent light being emitted by the sample. The higher the intensity of the light being emitted, the more concentrated the sample is with the target DNA. SYBR Green, as its name implies, fluoresces with a green color. Accordingly, it is important to use a different wavelength of light to excite the sample in order to prevent distorting the results. There are several types of fluorescent dyes similar to the SYBR Green used in this lab, but SYBR Green was chosen for its environmental friendliness and high level of light emission intensity to allow for easier detection of binding to the target DNA.
A fluorimeter is a device used to measure and identify fluorescence in a medium. Fluorimeters measure certain wavelengths of light and have two light detectors, one of which detects absorbency while the other detects fluorescence emission (So Dong 1). Single-Drop Fluorimeters do this with a single drop of a sample that is placed on a hydrophobic slide in the path of a beam of single wavelength light. For example, in our setup we used blue LED light to allow for light absorption and prevent contaminated intensity data that would occur if we used a green LED since the SYBR Green Dye fluoresces in the green range. A light detector or camera placed at a uniform distance from this sample is then used to capture the intensity of the fluorescence being emitted from the sample. Cameras are acceptable for this laboratory set-up even though they are unable to measure light intensity because it is assumed that light intensity is proportional to brightness; the brighter an object is in a photo, the higher the light intensity.
How the Fluorescence Technique Works
ProcedureSmart Phone Camera Settings
Placing Samples onto the Fluorimeter
Data AnalysisRepresentative Images of Samples
Image J Values for All Samples
For the sake of clarity, the samples are listed below with their corresponding Patient ID Number and corrected PCR Product Concentration. The Positive and Negative controls are also included.
Fitting a Straight Line
Compare each patient's results to the positive control value and the negative control value. Draw a final conclusion for each patient (positive or negative) and explain why you made that conclusion.
Works CitedZipper, Hubert, Herwig Brunner, Jurgen Bernhagen, and Frank Vitzthum. "Investigations on DNA Intercalation and Surface Binding by SYBR Green I, Its Structure Determination and Methodological Implications." Ncbi.nih.gov. US National Library of Medicine, 12 July 2004. Web. 3 Apr. 2014. So, Peter TC, and Chen Y. Dong. "Fluorescence Spectrophotometry." Scribd. Macmillan Publishers Ltd, Nature Publishing Group, 16 Jan. 2009. Web. 03 Apr. 2014. |