Descriptive Stats and Graph

Temperature:

Temp (°F) Gold Standard Spree
Average 96.64716049 95.5308642
Standard Deviation 1.922602071 0.870378299
Count 324 324
Standard Error 0.106811226 0.04835435



Heart Rate:

Heart Rate (bpm) Gold Standard Spree
Average 98.08976898 98.95379538
Standard Deviation 23.03054395 24.87753802
Count 303 303
Standard Error 1.32307018 1.429177217



Inferential Stats

Paired T-Test and Pearson's r were ran between the Golden Standard data and the Spree Band data for Temperature and Heart Rate.

Temperature:

Temp (°F) Data
T-Test 1.14066E-21
Standard Deviation 0.191688017



Heart Rate:

Heart Rate (bpm) Data
T-Test 0.427116193
Standard Deviation 0.690806489



Design Flaws and Recommendations

There were multiple 0 bpm readings recorded for both the gold standard and the Spree headband. Due to the data being from previous classes, it is unknown whether these zeroes were inserted due to missing data or if the machines had errors in their sensors, resulting in a reading of zero. For future research, missing data should be left blank instead of inserting a value of 0 in order to prevent inaccurate statistics being recorded.