Note: The heart rate was measured with a pulse oximeter. <br>
''Note: The heart rate was measured with a pulse oximeter.'' <br>


Gold Standard - Average: 96.4974026 <br>
Gold Standard - Average: 96.4974026 <br>
Line 51: Line 51:
[[Image:Temperature.jpg]]
[[Image:Temperature.jpg]]


Note: The temperature was measured with an oral thermometer. <br>
''Note: The temperature was measured with an oral thermometer.'' <br>


Gold Standard - Average: 95.48115152 <br>
Gold Standard - Average: 95.48115152 <br>

Revision as of 15:03, 24 September 2016

BME 100 Fall 2016 Home
People
Lab Write-Up 1 | Lab Write-Up 2 | Lab Write-Up 3
Lab Write-Up 4 | Lab Write-Up 5 | Lab Write-Up 6
Course Logistics For Instructors
Photos
Wiki Editing Help


OUR TEAM

Yesenia Barrera Millan
Mariam El Sheikha
Vishvak Rangarajan
Tori Johnson
Reese Morehouse

LAB 3 WRITE-UP

Descriptive Stats and Graph

Heart Rate

Note: The heart rate was measured with a pulse oximeter.

Gold Standard - Average: 96.4974026
Spree - Average: 97.65584416

Gold Standard - Standard Deviation: 25.99859918
Spree - Standard Deviation: 27.10634861

Pearson's R: 0.730784539

Temperature

Note: The temperature was measured with an oral thermometer.

Gold Standard - Average: 95.48115152
Spree - Average: 93.79393939

Gold Standard - Standard Deviation: 10.76273631
Spree - Standard Deviation: 12.81217932

Pearson's R: 0.797813158


Inferential Stats

Heart Rate

P-Value (Paired): 0.298225789

Temperature

P-Value (Paired): 9.25843E-05


Design Flaws and Recommendations

Based on the experimental results, it is determined that the Spree Band tracker is not effective for the intended purposes. The temperature and the heart rate were recorded from the headband, while at the same time they were recorded by the "gold standards": a thermometer as well as a pulse oximeter. After the results were recorded from all the devices and the Spree Bands were compared to the measurements taken from the gold standard devices, it was the decided that the Spree Band was not effective in the measurements, the device is flawed and provided inaccurate measurements, and is altogether unappealing to consumers.