LAB 3 WRITE-UP

Descriptive Stats and Graph

Temperature:

Averages:

- Gold Standard: 96.647 (°F)

- Spree Headband: 95.531 (°F)

Standard Deviations:

- Gold standard: 1.923

- Spree Headband: 0.870

Count:

- Gold Standard: 324

- Spree Headband: 324

Standard Error:

- Gold Standard:.107

- Spree Headband: .048

Two Tailed Paired T-Test:

p-value: 1.09E-21

Pearson's Correlation Coefficient:

r = 0.193


Heart Rate:

Averages:

- Gold Standard: 98.090

- Spree Headband: 98.954

Standard Deviations:

- Gold standard: 23.0305

- Spree Headband: 24.8775

Count:

- Gold Standard: 303

- Spree Headband: 303

Standard Error:

- Gold Standard: 1.323

- Spree Headband: 1.429

Two Tailed Paired T-Test:

p-value: .42712

Pearson's Correlation Coefficient:

r = 0.691




Inferential Stats

(Ashwin)





Design Flaws and Recommendations

(Nadiah)

Some of the design flaws of our device & the recommendations to overcome those include:


1. The application of ultrasound waves which pose negative effect on body tissues.

One is thermal effect where the ultrasound waves heat body tissues up. Another is cavitations, where small bubble of gas are released upon exposure to extreme negative pressure. Besides, high decibels of ultrasound waves would injured hearing.

Recommendation:


2. The need for a proper charging system without having to remove the bracelet from the patient

The bracelet shouldn't be removed at all from the patient even for charging because of the unforeseen heart attack.

Recommendation: A mini power bank should be included with the bracelet as a set. When it's time to charge the bracelet, it can be connected to the mini power bank via cable.


3. Allergies of certain patients towards plastic

The strap of our device would be composed of plastic material like polyurethane in order to make it durable and provide comfort to the patient. However, certain people are allergy towards plastic.

Recommendation: To be more flexible, we could also another version of this device, in which the strap is composed of latex.




Experimental Design of Own Device

-testing: people with heart conditions Select people that have a known risk for heart attack

How many?: (70 million people with high blood pressure) 1000 people 10 hospitals~ Throughout the US

People wear the device all the time and see how successful it is ^detect heart attacks and notification Did it detect the heart attack accurately or not

Detect blood pressure with two devices on same patient (blood pressure meter) ^paired T Test

Detect blood flow with a larger model ^paired T Test

Detect Heart Rate with finger heart rate sensor compared to wrist/patch heart rate sensor ^paired T test

As a patient goes to the hospital, their physical health will be measured on both devices