| + | * Optics: Double beam, chopper mirror | ||
| + | * Monochromator: Czerny-Turner monochromator, holographic grating (1600 lines/mm) | ||
| + | * Wavelength range: 190-900 nm, automated wavelength selection | ||
| + | * Slit width: 0.2 nm or 0.7 nm, manual selection | ||
| + | * Lamp turret: 2 lamps, simultaneously lit, manual selection | ||
| + | * Lamp mode: Emission, non-BGC, BGC-D<sub>2</sub> | ||
| + | * Frequency: 100 Hz | ||
| + | * Nebulizer: Nebulizer integrated impact bead and jacket tip, Pt/Ir capillary | ||
| + | * Chamber: Polypropylene | ||
| + | * Burner: Fixed position and height, 10 cm titanium slot burner, manual selection between air/C<sub>2</sub>H<sub>2</sub> and N<sub>2</sub>O/C<sub>2</sub>H<sub>2</sub> | ||
| + | * Gas control: Manual selection of flow rate, automatic air/N<sub>2</sub>O switching system | ||
| + | * Safety: Gas pressure monitoring, automatic flame monitoring, safety interlock for misuse of burner, automatic flame extinguish upon power failure | ||
| + | * Ignition: Manual selection | ||
| + | * Software: Based on Microsoft Windows, WizAArd | ||
| + | * Power requirements: AC220 V, 230 V, 50/60 Hz, 300 VA, requires transformer for AC115 V | ||
| + | * Ambient temperature range: 10~35°C | ||
| + | * Ambient humidity range: 45~80%, but less than 70% if temperature is above 30°C | ||
==References== | ==References== | ||
Revision as of 16:24, 13 November 2012
Contents |
Description
- Shimadzu AA-6200 double-beam atomic absorption spectrophotometer.
- It can detect the concentration of Au and Cu even at very low concentrations (ppm).
Purpose of Use
Location
Use
Safety
Specifications
- Optics: Double beam, chopper mirror
- Monochromator: Czerny-Turner monochromator, holographic grating (1600 lines/mm)
- Wavelength range: 190-900 nm, automated wavelength selection
- Slit width: 0.2 nm or 0.7 nm, manual selection
- Lamp turret: 2 lamps, simultaneously lit, manual selection
- Lamp mode: Emission, non-BGC, BGC-D2
- Frequency: 100 Hz
- Nebulizer: Nebulizer integrated impact bead and jacket tip, Pt/Ir capillary
- Chamber: Polypropylene
- Burner: Fixed position and height, 10 cm titanium slot burner, manual selection between air/C2H2 and N2O/C2H2
- Gas control: Manual selection of flow rate, automatic air/N2O switching system
- Safety: Gas pressure monitoring, automatic flame monitoring, safety interlock for misuse of burner, automatic flame extinguish upon power failure
- Ignition: Manual selection
- Software: Based on Microsoft Windows, WizAArd
- Power requirements: AC220 V, 230 V, 50/60 Hz, 300 VA, requires transformer for AC115 V
- Ambient temperature range: 10~35°C
- Ambient humidity range: 45~80%, but less than 70% if temperature is above 30°C